![Martel BETAGAUGE 330 Reference Manual Download Page 44](http://html1.mh-extra.com/html/martel/betagauge-330/betagauge-330_reference-manual_1732373044.webp)
14. Specifications
(15 °C to 35 °C unless otherwise noted.)
General
Instrument Setup Recall
5; last used on power-up
Environmental
Operating Temperature
-10 °C to +50 °C
Storage Temperature
-20 °C to +60 °C
Power Requirements
12 VDC
Battery
Eight (8) AA alkaline, Lithium or NiMh cells
Physical
Dimensions
8” H x 4” W x 2.375” D
Weight
2.5 lbs. (1.2 kg)
EMI/RFI Conformance
EN50082-1: 1992 and EN55022: 1994 Class B
Safety
IEC 1010 Low Voltage Directive
Connectors/Ports
Pressure - one, 1/8” NPT
BetaPort-P pressure module adapter;
RTD probe
Included Accessories
Soft case, batteries, manual, NIST-traceable certificate,
test leads and hose kit.
Ranges
-12 to 30.000 psi/psia
-12 to 150.00 psi/psia
-12 to 300.00 psi
mA
0 to 24.000 mA
Volts
0 to 30.000 VDC
RTD
-40.0°C to 105.0°C
(-40.0°F to 220.0°F)
Engineering Units
psi, bar, mbar, kPa, MPa, kgcm2, mmH
2
O @ 4°C, mmH
2
O @ 20°C,
cmH
2
O @ 4°C, cmH
2
O @ 20°C, inH
2
O @ 4°C, inH
2
O @ 20°C,
inH
2
O @ 60°F, mmHg @ 0°C, inHg @ 0°C
Accuracy
Pressure
All Ranges
6 months: .025% F.S.
1 year: .035% F.S.
mA
±0.015% of rdg ±0.002mA
Volts
±0.015% of reading ±0.002V
RTD (ohms)
±0.015% of rdg ±0.02 ohms; or ±0.1°C @ 0°C for Pt100
Temperature Effect (all functions)
No effect on accuracy on all functions from 15°C to 35°C
Add ±0.002% F.S./°C for temps outside of 15°C to 35°C
Optional LPT100A Probe
Meets PT-100 ALPHA 385 Class “A” Specifications
42