40
PT385_100;
PT392_100;
PTJIS_100;
CUSTOM;
For Example:
RTD_TYPE PT385_100 sets RTD type to PT100-385
RTD_TYPE
?
Returns the RTD type.
For Example:
If the RTD type is PT385_100, RTD_TYPE? Will return PT100_385.
SIM
Sets the output for current simulation. This command also switches the calibrator into
mA simulation mode. A number and a unit must be entered after the command.
For example:
SIM 5 MA sets the current simulation at 5 mA
SIM?
Returns the output of the current simulation. With the example above, the output
would be: 5.000000E-03, A
ST_START
Starts a switch test.
ST_CLOSE?
Returns the pressure that the switch closed at in the current pressure units.
ST_OPEN?
Returns the pressure that the switch opened at in the current pressure units.
ST_DEAD?
Returns deadband of the switch in the current pressure units.
TEMP_UNIT
This command is used to set the temperature unit used when measuring
temperature.
The first argument indicates which display to apply the change to. The second
argument is the unit, either CEL for Celsius or FAR for fahrenheit.
For example:
To set the temperature unit to fahrenheit on the lower display send TEMP_UNIT
LOWER, FAR.