background image

 

1.

 

Introduction

 

 

1.1 Description 
 

The Mark 25 Conductivity Analyzer is a simple, low cost, 

4 channel

, microprocessor-

based monitoring system designed for ultrapure water applications. The system 
consists of a readout module that can be used with up to four sensors; each sensor 
being incorporated into a flow-through style flow chamber. 
 
A large 4 digit LED display indicates either 

temperature

,

 raw conductivity

, or 

temperature-corrected conductivity

 while an 8-key membrane keypad allows the 

operator to quickly setup and calibrate the analyzer. 
 
The Mark 25 measures conductivity in 

three ranges

; 0-1 

µ

S/cm, 0-10 

µ

S/cm, and 0-

100 

µ

S/cm. Calibration is accomplished quickly via automatic calibration within the Mark 

25 or by comparison to a referee system (please refer to Martek’s ASTM & NIST-
traceable 

Mark 22 In Situ Calibrator

 for further information).  

 
Four alarm set points with relay contacts are standard in addition to four isolated analog 
recorder outputs (0-20 mA) 

and

 a serial RS232 digital output for hookup to external 

recording equipment. For maximum flexibility, 

alarm set points and analog outputs 

can be assigned to any range on any parameter on  any channel. 

 
Housed in a ¼  DIN corrosion-resistant plastic case, the readout module can be panel   
mounted with a minimum of tools. The analyzer accepts power inputs from 97 to 240 
VAC 50/60 Hz or 24 Vdc and comes with an 

ON/OFF switch

 located on the rear panel. 

Alarm and recorder wires are easily installed in removable terminal strips.  
 
Sensors connect via a DB-9 connector located on the rear panel of the Mark 25 and the 
sensors mount easily to any flat surface. In addition, Martek offers 

adapter plates

 

specifically designed to cover the space occupied by other brands of conductivity 
analyzers.  
 
Sensors come standard with 10 feet (3 meters) of cable.  However, longer cable lengths 
may be special-ordered from the factory up to lengths of 150 feet. 
 
 
 

NOTE: If the Mark 25 is purchased as a part of the Dissolved Carbon Dioxide 
Analyzer, the auto-ranging function is standard and cannot be changed.  For 
additional information, contact the factory. 

 
 
 
 

Summary of Contents for Mark 25

Page 1: ...Martek Instruments Inc Mark 25 Ultrapure Water Conductivity Analyzer Instruction Manual...

Page 2: ...rial number of the product and a brief description of the problem All shipments to MARTEK must be freight prepaid and addressed as follows MARTEK INSTRUMENTS INC 2609 Discovery Drive Raleigh NC 27616...

Page 3: ...ectrical connections to the monitor 6 3 Operation 3 1 Keyboard display 7 3 2 The CALIB Key 7 3 3 Calibrating Temperature 8 3 4 Calibrating Conductivity 9 3 5 Setting Alarms 9 3 6 Setting Recorder Outp...

Page 4: ...r outputs 0 20 mA and a serial RS232 digital output for hookup to external recording equipment For maximum flexibility alarm set points and analog outputs can be assigned to any range on any parameter...

Page 5: ...utput Alarm Output Four solid state relays 5 amp 250 V Weight 2 lbs 0 9 kg including one sensor with 10 feet 3 meters Construction Readout Module Case ABS plastic NEMA 12 IP65 Sensor Type 316 Stainles...

Page 6: ...Installation 2 1 Panel Cutout for Monitor 3 60 3 60 Panel Cut out 2 2 Sensor Dimensions 2 75 6 98 cm 1 75 4 44 cm Side View 8 00 20 32 cm 1 12 2 84 cm Top View 125 31 cm 196 49 cm 2 places 4 125 10 47...

Page 7: ...erly grounded outlet If recorder digital or alarm outputs are to be used make certain the Mark 25 is unplugged before connecting the cable to the appropriate recorder or computer Follow the wiring dia...

Page 8: ...ssed the Mark 25 will display rd If one of the bottom keys is selected the parameter high lighted in red will be displayed and its corresponding LED lit RANGE key Range changes can be made manually fo...

Page 9: ...rk 25 should be calibrated to ASTM or NIST traceable standards To calibrate temperature press the CALIB key The display will flash o C To select temperature press the UP ARROW key The flashing will st...

Page 10: ...y channel Press the CALIB until the display indicates the alarm prompt ALA Next press the UP ARROW key The display will prompt with AL 1 and all parameter LED s will be lit The UP ARROW key can be use...

Page 11: ...nction is ON the decimal point can be moved by pressing the RANGE key Press the CALIB key to select the desired parameter At this point the standard 4 20 mA output range can be changed to any value de...

Page 12: ...esired value has been reached press the CALIB key to set and activate the password Once the password is enabled the display will default to 1999 whenever the password prompt is selected To gain access...

Page 13: ...will be lit and flashing Use the UP or DOWN ARROW key to advance to the auto ranging prompt AUrG When the CALIB key is pressed the Mark 25 will display OFF To activate the auto ranging function press...

Reviews: