Martek Instruments Mark 25 Instruction Manual Download Page 9

 

  Calibrate temperature and 

conductivity 

 

 

  Assign and set alarm points 

 

 

  Assign and set recorder outputs 

 

 

  Set a password 

 

 

  Set up an analyzer ID, serial 

output, and auto-ranging 
functions 

 

 

Once the sensors have been connected 
to the Mark 25, the power cord can be 
plugged into any outlet supplying 97-240 
V AC 50/60 Hz power. When the Mark 
25 is switched on, it will immediately 
start monitoring temperature.  
 
Pressing the <CALIB> key will cause 
the Mark 25 to display a flashing 

o

C

.  

 
If you continue pressing the <DOWN 
ARROW> key the display will flash the 
following prompts: 

CON, ALA, rEC, 

PAS, uSr

. After 

uSr

, the Mark 25 

returns to monitoring temperature.  
 
Here are two 

basic rules

 to remember 

when using the <CALIB> key:

 

 

1.  Pressing the  <UP ARROW> key will 

select any flashing prompt. 

 

 

2.  Pressing the <RUN> key will return 

the Mark 25 to the monitoring mode. 

 
 

3.3 Calibrating Temperature

 

 
The Mark 25 can be calibrated 
automatically or manually for 
temperature. 
 

NOTE: 

Automatic, electronic calibration 

is a less accurate method of calibration. 
For best results, the Mark 25 should be 
calibrated to ASTM or NIST- traceable 
standards.

 

 
To calibrate temperature, press the 
<CALIB> key. The display will flash 

o

C

 

To select temperature, press the <UP 
ARROW> key. The flashing will stop. 
Next, press the <ZERO> key. The Mark 
25 will display three calibration bars and 
automatically zero the temperature 
before returning to the monitoring mode.  
 
To span temperature, press the 
<CALIB> key, then the <UP ARROW> 
key when the temperature prompt is 
displayed. Next, select the <SPAN> key. 
The Mark 25 will display 

Aut

 which is 

short for Automatic.  
 
To automatically calibrate temperature, 
press the <UP ARROW> key.  After the 
calibration bars appear, the Mark 25 will 
briefly display a span temperature of 

50.0

 before returning to monitor mode. 

 
The Mark 25 has now been calibrated 
for temperature. 
 
To manually span calibrate temperature, 
simply press the <DOWN ARROW> key 
when 

Aut

 is displayed. The Mark 25 will 

momentarily display three lines, then a 
number indicating the input voltage will 
appear in the display. Once this number 
has stabilized, press the <CALIB> key 
and select the appropriate span value 
using the up or down arrow keys (the 
display defaults to 25 deg. C).  When 
complete, press the <CALIB> key to 
enter the data into the Mark 25’s 
memory.  
 
 

Summary of Contents for Mark 25

Page 1: ...Martek Instruments Inc Mark 25 Ultrapure Water Conductivity Analyzer Instruction Manual...

Page 2: ...rial number of the product and a brief description of the problem All shipments to MARTEK must be freight prepaid and addressed as follows MARTEK INSTRUMENTS INC 2609 Discovery Drive Raleigh NC 27616...

Page 3: ...ectrical connections to the monitor 6 3 Operation 3 1 Keyboard display 7 3 2 The CALIB Key 7 3 3 Calibrating Temperature 8 3 4 Calibrating Conductivity 9 3 5 Setting Alarms 9 3 6 Setting Recorder Outp...

Page 4: ...r outputs 0 20 mA and a serial RS232 digital output for hookup to external recording equipment For maximum flexibility alarm set points and analog outputs can be assigned to any range on any parameter...

Page 5: ...utput Alarm Output Four solid state relays 5 amp 250 V Weight 2 lbs 0 9 kg including one sensor with 10 feet 3 meters Construction Readout Module Case ABS plastic NEMA 12 IP65 Sensor Type 316 Stainles...

Page 6: ...Installation 2 1 Panel Cutout for Monitor 3 60 3 60 Panel Cut out 2 2 Sensor Dimensions 2 75 6 98 cm 1 75 4 44 cm Side View 8 00 20 32 cm 1 12 2 84 cm Top View 125 31 cm 196 49 cm 2 places 4 125 10 47...

Page 7: ...erly grounded outlet If recorder digital or alarm outputs are to be used make certain the Mark 25 is unplugged before connecting the cable to the appropriate recorder or computer Follow the wiring dia...

Page 8: ...ssed the Mark 25 will display rd If one of the bottom keys is selected the parameter high lighted in red will be displayed and its corresponding LED lit RANGE key Range changes can be made manually fo...

Page 9: ...rk 25 should be calibrated to ASTM or NIST traceable standards To calibrate temperature press the CALIB key The display will flash o C To select temperature press the UP ARROW key The flashing will st...

Page 10: ...y channel Press the CALIB until the display indicates the alarm prompt ALA Next press the UP ARROW key The display will prompt with AL 1 and all parameter LED s will be lit The UP ARROW key can be use...

Page 11: ...nction is ON the decimal point can be moved by pressing the RANGE key Press the CALIB key to select the desired parameter At this point the standard 4 20 mA output range can be changed to any value de...

Page 12: ...esired value has been reached press the CALIB key to set and activate the password Once the password is enabled the display will default to 1999 whenever the password prompt is selected To gain access...

Page 13: ...will be lit and flashing Use the UP or DOWN ARROW key to advance to the auto ranging prompt AUrG When the CALIB key is pressed the Mark 25 will display OFF To activate the auto ranging function press...

Reviews: