3.3.2 ANGLE SENSOR CALIBRATION
From the LOADCELL INTERFACES menu move the cursor to ANGLE SENSOR and press enter.
Important: Before calibrating the angle sensor ensure the machine is level.
Use the key to select the orientation of the loadcell
interface relative to the direction of travel, and then press
enter. (fig 13)
The interface icon on the screen indicates the orientation
of the interface when looking from the top down, as
compared to the direction of travel
To set the angle sensor Zero Point (TARE) press the TARE
key. (fig 14)
Ensure the spreader is level, and press enter.
The Angle Sensor calibration is then complete. (fig 15)
3.3.3 FACTORS STORE
For extra security calibration data can be manually saved to a separate memory area called
“STORE A”. This data will NOT be over written each time the instrument is turned off, but it WILL
be cleared after a Total Software Reset. (RESET FACTORS)
From the FACTORY CONFIG menu select FACTORS STORE. (fig 16)
On the FACTORS STORE MENU move the cursor to SAVE
to store A, press enter to save all factors.
To RESTORE factors move the cursor to Restore from A,
and press enter.
It is a good idea to regularly store calibration factors to
STORE A to prevent unwanted data loss.
pg 15
fig 13 : Angle Calibration Screen 1
fig 14 : Angle Calibration Screen 2
fig 15 : Angle Calibration Screen 3
fig 16 : Factors Store Screen