DEGASSING SYSTEM DG-50
INSTALLATION OPERATION & MAINTENANCE MANUAL
MA-DG-50r01
13
Degassing system electric wiring diagram
5
START UP
DANGER
:
TO AVOID THE RISK OF BURNS, IT IS ESSENTIAL BEFORE START-UP, VERIFY THAT IS
NOT POSSIBLE TO ACCESS THE REAR PART OF THE MOUNTING PLATE. IF INSTALLED AS SPECIFIED
IN THIS MANUAL, THERE IS NO RISK FOR ACCIDENTAL CONTACT WITH HOT PARTS. FOLLOW THIS
PROCEDURE FOR SAFE OPERATION:
•
Open the cooling water inlet and outlet valves (if installed), for non self-cooling systems and
ensure that a full flow can be seen at the cooling water outlet.
•
Gradually open the sample inlet valve (not included in the system) and regulate the flow to
achieve a value between 100 to 250 cm³/min. Then is necessary to wait the time required for
the sample to go out from sample outlet.