Mantracourt Electronics Limited
T24-PR1 User Manual
10
Input Settings
Here you can set the action to take when the switch contacts are closed, set the Data Tag that will trigger an
output and also set the Data Tags of the data used as the inputs along with how the data is formatted.
Items you can change:
Switch Mode
NOTE: The external button/switch option will be made available on future releases of the
T24-PR1
Click to select either:
– Triggers an output from the serial port. Note that
Min Interval
setting
on the Output Settings page may stop this from working at the rate the
contact closes.
Gross/Zero
– Toggles the measurement mode between gross and net. When
switching to net the net value is zeroed (tared). This will affect the value of
the <N> token which is the net value of all summed inputs.
Output Trigger Data Tag
Enter the Data Tag which, on arrival, will trigger an output (I.e print). Note
that
Min Interval
setting on the Output Settings page may stop this from
working at the rate at which the data arrives.
This is usually set to the Data Tag of one of the inputs.
Example:
If the output was being sent to a serial display the same Data Tag
would be assigned to input 1 and the output trigger so that each time data
arrived the output would be sent to the serial display.
Remote data Tags and Timeouts
Data Tag
Enter the Data Tag (in hexadecimal) to supply data to this input.
NOTE: You can click the P button to retrieve the Data Tag of a device by