4.
Insert the syringe with a narrow tip into the back, left hole as shown, the tip should fit snuggly into the
hole so that no air is introduced.
5.
Select MENU/ OPERATION / PRIME LINES / PRIME PORT A
6.
Press ENTER, and at the same time gently push on the plunger of the syringe. You must prime Port A to
open the valves of the internal fluidics path. Observe the flow of water out of the Port A tubing and
note any particles that are expelled into the waste beaker. Stop pushing the plunger when you hear the
pump stop.
7.
Repeat Step 6 a few times and ensure the flow out of Port A is strong.
8.
Repeat the same process for Port B, using the same syringe tip and hole position.
9.
Prime as normal to verify that the blockage has been removed.