12
DATA
LEVEL 1
DESCRIPTION
Result Log
Data display. Presents result data for previous samples.
Use the Up, Down, Left & Right arrow touch keys to move within each individual log.
Event Log
Data display. Presents event data for previous samples.
Use the Up, Down, Left & Right arrow touch keys to move within each individual log
Calibration Log
Data display. Presents calibration data for previous samples.
Use the Up, Down, Left & Right arrow touch keys to move within each individual log
OPERATION
LEVEL 1
LEVEL 2
DESCRIPTION
Prime Lines
Prime Port A
Primes Port A.
Prime Port B
Primes Port B.
*Check Seals
Checks that the PeCOD
®
fluidics are sealed.
Run Sample
Commences operation of the analyser.
Run Calibration
Commences the calibration routine.
Run Standby
Commences standby operation which primes Port B using a repeat
cycle time as specified.
*Run Wash
Commences the wash routine.
*Open Analyser
Purges line to allow analyser to be opened and sensor to be
removed.
*
Feature which is not implemented in current L100 practises
SET UP
LEVEL 1
LEVEL 2
LEVEL 3
DESCRIPTION
Analysis
Method
COD Range
Choice of:
Advanced Blue (<20 ppm analysis);
*Blue (<20 ppm analysis);
Green (<150 ppm analysis)
Yellow (<1,500 ppm analysis);
Red (<15,000 ppm analysis)
Use the Up & Down arrow touch keys to select
COD range.
*Blue range is not currently implemented in L100
practises
*Set Algorithm
Normal or Pulsed.
Normal method is continuous oxidation. This is
the default method.