14
LEVEL 1
LEVEL 2
LEVEL 3
DESCRIPTION
Periodic
Actions
*Cal. Interval
Cal. Cycles.
Not for use for L100
*Wash Interval
Wash Cycles.
Not for use for L100
Standby Interval
Standby (HH:MM)
Default value set to 00:00 (hours: minutes).
Change value to set standby mode interval time.
Use the Up & Down arrow touch keys to change
value.
Reference
Solution
Reference
Input to set calibrant concentration. Valid values
are from 0.0 to 99999.9. Note: this is the
concentration prior to dilution with electrolyte.
Use the Up & Down arrow touch keys to set the
value.
Set Date and
Time
YYYY/MM/DD
HH:MM:SS
Input to set up date and time or to change.
Use the Up & Down arrow keys to set date/time.
Display Units
COD Units
COD units.
Choice of “ppm or mg/L”.
Use the Up & Down arrow touch keys to choose
display units.
Equiv. BOD Units
BOD Factor.
If set to Zero then no BOD data is displayed.
Minimum value: 0.00
Maximum value: 9.99
Use the Up & Down arrow touch keys to enter
COD to BOD multiplication factor.
*
Feature which is not implemented in current L100 practises
DIAGNOSTICS
LEVEL 1
LEVEL 2
DESCRIPTION
Factory Settings
Reset to factory
settings.
Resets all parameters to factory settings (default values).
Select YES or NO.
System Reset
Restarts the analyzer.
Set Pin Number
Default PIN = 0000
In this setting, security is switched to OFF.
Use the Up & Down arrow touch keys to change value.
Erase Logs
Erase Result Log
Erases result log.
Erase Event Log
Erases event log.
Erase Cal. Log
Erases calibration log.