Trouble shooting guide

Compressor  does  not  run  when  ‘up’  is 

selected on the 2-way hand control.
Ensure that the air hose from the product 

being inflated is connected to the Airflo.

Compressor keeps stopping when trying 

to inflate the lifting aid.

Check  that  there  are  no  kinks  in  the  air 

hose.

Airflo 12

Weight: 

3.7Kg  (8 lbs)

Output pressure: 

0.25 bar  (3.75 psi)

Batteries: 

1 x 12v - 2200mAh Ni-MH battery pack

Battery Life: 

Greater than 500 charging cycles (2 - 5 years depending on usage)

Environmental protection:

 

(Airflo 12 housing)

 

IP51 (Protection against large dust deposits & vertical falling drops  

   

of water)

 

(Push-button hand control)

  IP67 (Totally protected against dust & immersion in water)

T

ype BF equipment

This equipment has been tested and meets the requirements of current EMC legislation

Charger: 

Model number

 

- HX1506SR

Input: 

AC 100v  - 240v,  50-60Hz, 0.5A Max

Output: 

DC 15v, 0.35A

Class II Double insulated equipment

Check  that  there  are  no  kinks  in  the  air 

hose.

This  happens  when  the  compressor’s 

internal temperature cut-out has tripped. 

Leave the Airflo

 

12 for 60-90 minutes to 

cool.

Bottom  Battery  Status  Indicator  light 

flashing ‘Red’.

Compressor causes electrical 

interference with another piece of 

electrical equipment.

Move  either  the  compressor  or  other 

item further away from each other until 

the interference stops.

Specifications

Summary of Contents for Airflo 12

Page 1: ...Simple solutions for everyday independence English User Instructions Airflo 12...

Page 2: ...leading to the potential for fire or even an explosion in the presence of a spark Ensure that when in use the Airflo 12 and its connected hose and hand control lead are laid out in such a way as to a...

Page 3: ...lift When the compressor stops running the battery status indicator will show a bottom green light this will be red if the battery charge has become low To determine the true battery state following a...

Page 4: ...ed DO NOT USE Fit the appropriate mains plug adaptor to the battery charger A Plug the battery charger into a mains power socket and connect the charger output lead to the Airflo 12 charging socket B...

Page 5: ...p or Down buttons are pressed the internal microprocessor carries out a battery voltage check and if the voltage becomes too low the Down button will not work This feature will always ensure that the...

Page 6: ...ted as follows Unplug the Airflo 12 from the battery charger Clean the external surfaces using a non abrasive damp cloth with warm water Wipe dry thoroughly with a dry towel before reconnecting to the...

Page 7: ...ainst large dust deposits vertical falling drops of water Push button hand control IP67 Totally protected against dust immersion in water Type BF equipment This equipment has been tested and meets the...

Page 8: ...g Direct current DC Type BF equipment protection against electric shock Max Min operating environment temperature Attention consult instructions Warning Class II Double Insulated equipment Dangerous v...

Page 9: ...warranty period the defect will be rectified free of charge providing that a Reasonable evidence is provided e g purchase invoice delivery note that the claim is being made within the warranty period...

Page 10: ...lectrical Equipment General requirements for basic safety and essential performance IEC 60601 1 11 Medical Electrical Equipment Requirements for use in the home care environment IEC 60601 1 2 Medical...

Page 11: ...Simple solutions for everyday independence Airflo 12 11 Please note that these instructions are available in other languages and can be downloaded from our website by visiting www mangarhealth com...

Page 12: ...ns for everyday independence MI0353 issue 1 04 17 Mangar International Ltd Presteigne Powys Wales LD8 2UF United Kingdom Tel 44 0 1544 267674 Fax 44 0 1544 260287 E mail sales mangarhealth com www man...

Reviews: