Airflo
12
4
2
First 2
Hours
16
Hours
NOTE:
The charger is not capable of providing sufficient power to charge the batteries while the
compressor is running.
To maintain battery condition, it is recommended that the Airflo
12
is left continuously on charge
when not in use. The batteries cannot be overcharged using this charger.
WARNING:
DO NOT recharge the batteries in a bathroom or in a damp environment.
To disconnect the charger from the mains power supply, unplug the charger from the mains power
socket.
Battery Charging
- figs.
2
•
Check the battery charger casing and lead
for signs of damage and if damaged DO NOT
USE.
•
Fit the appropriate mains plug adaptor to
the battery charger (
A
).
•
Plug the battery charger into a mains power
socket and connect the charger output lead
to the Airflo
12 charging socket (
B
).
NOTE:
The batteries must be recharged for a
minimum of 16 hours to achieve full capacity.
When the battery is being charged:
•
For the first 2 hours, the bottom light will
show a steady red and the next light will be
flashing green. After 2 hours the lights will
change to two steady green and one flashing
green. After 16 hours all lights will become
steady green.
NOTE:
If charging is momentarily interrupted
(a power cut, switching the power socket off
or disconnecting the charging lead), on re-
connection, the charging process will start at
the beginning and follow the sequence above
for the next 16 hours.
B
A