www.mc-techgroup.com
SP30-H../EX2 | 1.02.00
13
13.1
ELECTRICAL CONNECTION OF THE HEATED SAMPLE TUBE
W A R N I N G !
The used cables for the connection of the heating must have an
outer diameter of 6-12 mm corresponding to the clamping range of
the cable glands!
W A R N I N G !
The max. admissible ambient temperature for the connection box
of the sample tubes SP30-H../EX2 and SP30-H1.1-V/EX2 is +70 °C
[158 °F].
Temperature resistant cable have to be used corresponding to the
max. temperature at the installation site.
•
Remove lid of the connection box. In the lid there is also the electrical wiring plan.
•
Insert the power cable (min. 3 x 1.5 mm
2
) coming from the external temperature controller
through the cable gland and connect it at the corresponding terminals 6, 7 and 8.
•
Insert the temperature sensor cable through the other cable gland and connect it to the terminals
2, 3 and 4.
•
Remount lid.
Function and correctness of the temperature control (temperature
class) is in the sole responsibility of the installer and operator.
Valid from serial number 1810XXXX