9.
Errors and solutions
The protective diaphragm effectively prevents that liquids or solids are sucked into the
pump. If substances are sucked in nevertheless or any other problem occurs, immediately
put the device out of operation. Contact your dealer or the customer service of your shop.
The breast pump/power unit must not be opened by users/patients. The device is
maintenance-free. For repair work, the device must be returned to the manufacturer.
10. Environmental protection and disposal
Please dispose of a disused
mamivac
®
LACTIVE
breast pump in accordance with your local
disposal regulations for electrical devices.
11. Technical data
The expected service life is 2 years.
mamivac
LACTIVE
REF 281024
Nominal voltage:
12 V DC
Power consumption:
1.00 A
H x W x D (mm):
115 x 125 x 195
Weight:
630 g
Max. vacuum:
-33 kPa (± 15%)
Mean vacuum:
Device for intermittent sucking without continuous vacuum
Power supply unit
REF 283005
Nominal input voltage:
100-240 V AC 50 / 60 Hz
Output voltage:
12 V DC
Protection class:
II
12. Conversion table of units
1 kPa
corresponds:
0.01 bar
0.14504 PSI
corresponds:
7.5006 mmHg
(pounds per square inch)
(millimetre mercury, at 0°C)
- 23 -
Device classification:
Classified acc. to 93/42/EEC:
Article 9, Appendix IX:
Class IIa/Rule 11
Problem
Error/solution
Pump does not start
- Check connections of power supply.
- LED on power unit must light (if existing).
- If the problem is not solved, please contact the customer
service of your shop.
Pump does not suck
(does not build up
vacuum)
- Switch off and on the device several times.
Wait a minute before switching on again.
-
Check the respective suction kit for proper assembly.
- Check tube valve (for cracks and dirt).
- Check protective diaphragm for cracks.
- Protective diaphragm must be inserted flush and crinkle free.
- Check hose connections.
Summary of Contents for Lactive 287000
Page 95: ...REF 287000 2020 05 GB1030009RU...
Page 98: ...mamivac 1 3 2 mamivac LACTIVE REF 281024 a Vacuum b 2 Phase c d e 98 d c e a b g h i j f...
Page 99: ...f g h i j 3 mamivac 4 5 5 2 5 1 2 mamivac mamivac LACTIVE 3 1 1 60 2 99...
Page 100: ...3 2 2 30 3 3 4 4 1 4 2 mamivac 100 C 15 mamivac 100...
Page 103: ...7 12 300 12 300 8 mamivac mamivac LACTIVE 3 125 100 C o 5 3 2 10 103...
Page 104: ...8 1 16 29 C 3 4 4 C 3 18 C 6 2 3 mamivac REF 282100 9 10 mamivac LACTIVE 104...
Page 107: ...287000 2020 05 GB1030009GR...
Page 111: ...3 mamivac 4 5 5 mm 2 5 mm 1 m 2 mamivac mamivac LACTIVE 3 1 1 60 2 3 2 2 30 111...
Page 112: ...3 3 4 4 1 4 2 mamivac 100 C 15 mamivac 50 5 112...
Page 115: ...mamivac LACTIVE 3 125ml 100ml 5 3 2 10 8 1 16 29 C 3 4 4 C 3 18 C 6 2 3 mamivac 282100 115...
Page 119: ...a REF 287000 2020 05 GB1030009UA...
Page 123: ...3 mamivac 4 5 5 2 5 1 2 mamivac mamivac LACTIVE 3 1 1 60 2 3 2 2 30 123...
Page 124: ...3 3 4 4 1 4 2 mamivac 100 C 15 mamivac 50 124...
Page 125: ...5 mamivac LACTIVE mamivac LACTIVE 5 1 EM mamivac LACTIVE 11 14 mamivac LACTIVE 125...
Page 127: ...12 300 8 mamivac mamivac LACTIVE 3 125 100 5 3 2 10 127...
Page 128: ...8 1 16 29 C 3 4 4 C 3 18 C 6 2 3 mamivac REF 282100 9 10 mamivac LACTIVE 11 2 128...
Page 131: ...131...
Page 132: ......