Chapter 4
System Overview
Page 4-18
MAN 0486
ill 8485
g.
Check again for bubbles in the cell. Tap the cell gently to dislodge these.
h.
Ensure that both electrodes on the inside of the cell are completely immersed
in the sample.
i.
Remove the syringe and insert a cell stopper in each port:
ill 7940
j.
Remove any liquid spilt on the outer part of the electrodes and the outside of
the cell.
Load the cell
Depending on the test being run, one of these prompts is displayed:
For a
Size test
: