Operation Manual PILOT TEST
8504660
Rev.
3
41
06/11/2019
Recommended separation distances between portable and mobile RF communications equipment and the PILOT TEST.
The PILOT TEST is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the PILOT TEST can help prevent electromagnetic interferences by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the PILOT TEST as recommended below, according
to the maximum output power of the communications equipment.
Rated Maximum output
power of transmitter
[W]
Separation distance according to frequency of transmitter
[m]
150 kHz to 80 MHz
d
= 1.17
√𝑃
80 MHz to 800 MHz
d
= 1.17
√𝑃
800 MHz to 2.7 GHz
d
= 2.23
√𝑃
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.70
3.70
7.37
100
11.70
11.70
23.30
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where
P
is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
Note 1
At 80 MHz and 800 MHZ, the higher frequency range applies.
Note 2
These guidelines may not apply to all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The PILOT TEST is intended for use in the electromagnetic environment specified below. The customer or the user of the PILOT
TEST should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test
level
Compliance
Electromagnetic
Environment-Guidance
Electrostatic Discharge
(ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be greater than 30%.
Electrical fast
transient/burst
IEC61000-4-4
± 2 kV for power supply lines
100 kHz repetition frequency
± 1 kV Line-to-line
100 kHz repetition frequency
± 2 kV
± 1 kV
Mains power quality should be that of
a typical commercial or residential
environment.
Surge
IEC 61000-4-5
± 1 kV Line-to-line
± 2 kV Line-to-ground
± 1 kV
± 2 kV
Mains power quality should be that of
a typical commercial or residential
environment.
Voltage dips, short
interruptions and voltage
variations on power
supply lines
IEC 61000-4-11
0%
U
T for 0.5 cycle
0 % UT for 1 cycle
and
70%
U
T
for 25/30 cycles
Single phase: at 0°
0%
U
T for 0.5 cycle
0 % UT for 1 cycle
and
70%
U
T
for 25/30 cycles
Single phase: at 0°
Mains power quality should be that of
a typical commercial or residential
environment. If the user of the PILOT
TEST requires continued operation
during power mains interruptions, it is
recommended that the PILOT TEST
be powered from an uninterruptable
power supply or its battery.
Power frequency (50/60
Hz)
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or residential environment.
Note: U
T is the A.C. mains voltage prior to application of the test level.