![Maico PILOT TEST Operation Manual Download Page 37](http://html1.mh-extra.com/html/maico/pilot-test/pilot-test_operation-manual_3713140037.webp)
Operation Manual PILOT TEST
8504660
Rev.
3
35
06/11/2019
DEVICE SPECIFICATIONS
Environmental
conditions:
Operation: +15 °C to +35 °C /
+ 59 °F to +95 °F
Relative humidity 30 % to 90 % (non-condensing)
Air pressure 98 kPa to 104 kPa
1
Maximum altitude: 2000 m / 6561 ft above sea level
Warm up time: approx. 1 minute (incl. boot up time)
Storage:
0 °C to + 50 °C / 32 °F to +122 °F
Humidity 10 to 95 % (non-condensing)
Transport: -20 °C to + 50 °C / -4 °F to +122 °F
Humidity 10 % to 95 % (non-condensing)
Weight:
1.3 kg / 2.9 lbs
Dimensions:
305 mm x 260 mm x 65 mm (12.01 in x 10.24 in x 2.56 in)
Display:
TFT LCD Panel 5.0 in
Active size: 110.88 mm x 62.83 mm (4.37 in x 2.30 in)
User interface:
Push button
Language Settings
English, French, German, Spanish
1
Environmental conditions during operating according IEC 60645-1.
NOTE
: Reference equivalent threshold sound pressure levels may differ significantly with ambient
pressures outside the above range. Therefore recalibration around the normal ambient pressure at
the site of the user should be undertaken in those circumstances where the calibration site and the
user site do not share similar ambient conditions.