To check the calibration data for the pickup type
in use:
1. Open the "Pickup" menu by pressing the
key.
The information boxes "Last cal." and "Next
cal." display the most recent and the next
calibration to be carried out.
2. To modify the interval between the most
recent and next calibration:
Activate the "Interval" button by press
•
ing the key and enter the desired inter
val (in days).
The date for the next calibration will be
modified automatically.
Some of the functions of the "Pickup"
menu can only be accessed after enter
ing the code number.
30
To print the list of pickup types and their correc
tion values:
1. Display the "Pickup" menu, then press the
key.
2. In the "Print" menu select the "Previous
view" button by pressing the key.
If the calibration interval has passed
for a particular pickup or if it has
not yet be calibrated, this is indi
cat ed with an asterisk follow
ing the meas uring range, e. g.
"T0 Pxx350 350* 0".
30 If the code number "00000" set at the factory is used, the
"Pickup" menu is freely accessible.
i
i
6
Calibrating and Setting the Pick-up Correction Value
The sensitivity of Mahr pickups is adjusted at the
factory so they can be used for meas urements
straight away.
Sensitivitymatched pickups are set to their
nom inal sensitivity. If the value 0.0 % is displayed
in the "Correction" input box in the "Pickup"
menu, the pickup used will operate at nominal
sensitivity.
If necessary, to reach a higher degree of preci
sion, pickups can be calibrated
29
. Dynamic
calibration determines the deviation of a pickup
from its nominal sensitivity. The instrument calcu
lates a correction value to be taken into consider
ation in the evaluation.
This allows deviations lying within the tolerance
for the entire meas uring station (±15 %) to be
compensated.
For calibration,
– a separate roughness standard
(e. g PRN 10)
or a separate geometric standard
(e. g. PGN3 )
– or the integrated standard (only for the
PHT 6350 standard pickup)
can be used.
If the integrated standard is used for cali
bration, it should be checked at regular
intervals using a calibrated skidless pick
up to ensure that the given
Rz
nominal
value is still valid (see Section 9.1).
If the
symbol is displayed in the top
status bar, the selected pickup type has
not yet been calibrated or the calibration
interval has passed.
29 The instrument settings cannot be modified if they are
blocked (see Section 3.5.2).