Mahr GmbH, MarSurf M 300 and RD 18
21
In special cases, the next shortest cutoff may
be used instead of the standard one required
by ISO 4288. The "Lc short" setting can also be
implemented after the meas urement.
14
After the setting is made, the cutoff value is dis
played in the top status bar
(5) (see Fig. 2)
.
When the short cutoff is switched on, the symbol
appears in the top status bar.
If a standard traversing length is selected
by pressing
Lt/Lc
, the short cutoff is de
activated again.
Periodic
profiles
Aperiodic profiles
Cutoff
Sampling length
Evaluation length
RSm
in mm
Rz
in µm
Ra
in µm
Lc
in mm
lr
in mm
lm
in mm
(with
n
= 5)
RSm
≤
0.13
Rz
≤
0.5
Ra
≤
0.1
0.25
0.25
1.25
0.13 <
RSm
≤
0.4
0.5 <
Rz
≤
10
0.1 <
Ra
≤
2
0.8
0.8
4
0.4 <
RSm
10 <
Rz
2 <
Ra
2.5
2.5
12.5
Tab. 2
Determining the cutoff
in accordance with DIN EN ISO 4288
14 If the automatic setting
A
of cutoff and traversing length
is set, the cutoff cannot be modified.
i
Selecting the length of the pre- and post-
travel
If standard traverse lengths cannot be used
with the testpiece, the pretravel
Lv
and post
travel
Ln
can be shortened
15
. In the setting
"Lv short"
Lv
and
Ln
are half as long as the stan
dard cutoff.
When the short pre and posttravel are switched
on, the symbol
appears in the top status
bar.
By selecting a standard traversing length
with
Lt/Lc
the short pre and posttravel
are switched off again.
15 If the automatic setting
A
of cutoff and traversing length
is set, the length of the pre and posttravel cannot be
modified.
i