6. OPERATION 
 

6.1 POWERING UP THE MSV-2010 

 

To operate the MSV-2010: 

1.Attach the endoscope to the opto-mechanical adapter connected to the camera head  (refer to Section 5.5). 

2. 

Attach the light source to the endoscope. 

3.  Turn on the video camera, monitor and light source. Wait until SETTING list is appeared on the monitor 

screen. Do not press any front panel buttons or camera head buttons and any keyboard key until date, time and 
patient data window are appeared on the monitor screen. 

4.  Use the focus ring of the opto-mechanical adapter to focus the image. The image positioning on the screen can be 

adjusted by the free-spin/orientation mechanism  (refer to Section 6.2) and the optional centering mechanism (refer 
to Section 6.3), of the opto-mechanical adapter. 

5. 

Perform the white balance procedure (refer to Section 6.5.1). 

6.   At any time the image can be adjusted using the touch-buttons on the front panel of the camera control unit 

by: 

      a) Adjusting the brightness level (refer to Section 6.5.3). 

      b) Using gain control (refer to Section 6.5.2). 

 

NOTE 

- When the image is being adjusted, text messages appear on the upper line of the message 

  

 

window of the monitor screen display. 

 
 

6.2 FREE-SPIN/ORIENTATION MECHANISM 

 

1.  While using a flexible endoscope, release the free-spin lock of the opto-mechanical adapter (refer to Figure 3) 

and rotate the endoscope until the arrow in the view field points to twelve o’clock. Then return the free-spin 
lock to its locked position. 

2.  When using rigid endoscopes, release the free-spin lock of the opto-mechanical adapter, rotate the endoscope 

to the required position, then return the free-spin lock to its locked position.  

 

 

6.3 CENTERING MECHANISM 

1.  When the opto-mechanical adapter has a centering mechanism (optional), it can be used to center the image 

from a flexible endoscope on the screen.  

2.  The centering mechanism has two adjustment knobs (refer to Figure 3). When the view field of the flexible 

endoscope is correctly oriented at twelve o’clock, the knobs are positioned on the top and the right side of the 
mechanism. Turn these knobs to move the image up or down (right knob) and right or left (top knob) on the 
monitor. 

 
 
 
 

Summary of Contents for MSV-2010

Page 1: ...MSV 2010 Digital Video Camera Operator Manual...

Page 2: ......

Page 3: ...OLS 21 6 5 NORMAL OPERATION MODE 22 6 5 1 White Balance 22 6 5 2 Gain Control 22 6 5 3 Brightness Adjustment 22 6 6 MENU OPERATION MODE 23 6 6 1 Main Menu 23 6 6 2 Picture Setting Sub Menu 23 6 6 2 1...

Page 4: ...6 8 MENU SUMMARY 32 7 CLEANING 33 7 1 CAMERA CONTROL UNIT 33 7 2 CAMERA HEAD 33 7 3 OPTO MECHANICAL ADAPTER 33 7 4 GLASS WINDOW ON CAMERA HEAD AND OPTO MECHANICAL ADPTER 33 8 DEZINFRCTION AND STERILIZ...

Page 5: ...a live picture in picture PIP image Capability to adjust the window used for automatic shutter control Capability to manually change shutter speed Capability to adjust image contrast Enhancement Capa...

Page 6: ...ces do not exceed the capacity of the isolation transformer For further information contact your local MAHE distributor WARNING Never drop this equipment or subject it to severe impact as it could com...

Page 7: ...ic shock do not open or expose the camera control unit to rain or moisture Refer all servicing to qualified personnel only 2 5 AFTER USE 1 Refer to the operating instructions supplied with all the com...

Page 8: ...plied part Safety degree specified by IEC 601 1 OFF Power disconnection from the mains ON Power connection to the mains Attention consult accompanying documents Equipotentiality CONVENTIONS The follow...

Page 9: ...1 0V C Burst 0 3V ptp 75 Ohm RGBS R G B 0 7V Sync 2 1V ptp 75 Ohm Power Supply 100 240VAC 50 60Hz automatic Power Consumption Approximately 20 Watt Optical Interface CS or C mount C mount with extensi...

Page 10: ...l Left scroll button in Menu mode 3 Green LED indicating activated Gain function 4 Brightness up button for manual brightness adjustment Up scroll button in Menu mode 5 Menu button for entering in Men...

Page 11: ...Keyboard connector 6 pin mini DIN type connector 6 Remote control connector 3 5mm jack to operate peripheral devices 7 Y C video outputs 2 x 4 pin mini DIN type connector 8 Composite video output BNC...

Page 12: ...12 4 1 3 Camera Head and Opto mechanical Adapter Figure 3 Camera head and opto mechanical adapter 1 3 4 5 2 6 Camera head Opto mechanical adapter 12 3 4...

Page 13: ...No Description 1 1 button refer to Section 6 14 2 2 button refer to Section 6 14 3 Free spin lock 4 Focus ring 5 Right left adjustment knob optional 6 Up Down adjustment knob optional...

Page 14: ...keys Include keys for letters numbers and special characters brackets commas etc 3 Num Lock key Locks function of number keys to print numbers only Status indicator will light when selected 4 Backspac...

Page 15: ...or special character pictured on the key to appear on the screen 8 Delete key Deletes the characters 9 Space Bar Creates space and deletes a previously typed character 10 Caps Lock key Caps Lock indic...

Page 16: ...een Main Screen Endoscopic image 4 User User Window Alpha numeric two lines field up to full line long 5 Message Message Window Alpha numeric one line field 16 characters long 6 Patient Data Name Alph...

Page 17: ...tion It may cause unrecoverable damage to the camera head To connect the camera head to the camera control unit align the red dot on the camera cable connector with the red dot on the CAMERA jack loca...

Page 18: ...deo Cable BNC BNC 1 5m 1 H10 0039 5 4 CONNECTING POWER 1 Connect the AC power cable to the inlet located on the rear panel of the camera control unit 2 Plug the AC power cable into a wall outlet 5 5 A...

Page 19: ...fer to Section 6 5 3 b Using gain control refer to Section 6 5 2 NOTE When the image is being adjusted text messages appear on the upper line of the message window of the monitor screen display 6 2 FR...

Page 20: ...b menu items shall be performed by pressing the and buttons The selected item shall appear on a green background Holding the or button for longer than approximately 0 5 second reference value shall sc...

Page 21: ...d there is no need to adjust it again unless the illumination conditions change 6 5 2 Gain Control In case of a dark or badly illuminated object the user can activate the gain control to further incre...

Page 22: ...MENU PICTURE SETTING SCREEN SETTING BUTTON SETTING CLOCK SETTING DEFAULT SETTING 2 To activate a new menu item select it by buttons or keys The selected item shall appear on a green background 3 To op...

Page 23: ...panel The message RED BLUE 20 to RED BLUE 20 will be appeared on the monitor screen indicating the color level Each pressing by the key or buttons after the maximum or minimum color level has been re...

Page 24: ...ing to image and display current setting To leave sub menu press Esc key or MENU button 6 6 2 7 Freeze Mode The FREEZE MODE function is allowed to user select the best quality of the frozen picture If...

Page 25: ...ton 6 6 3 5 Copy Mode B COPY MODE B function is intended to select the copy screen mode In COPY SCREEN mode the current picture is captured by video printer In ADD DATA mode the current picture is cap...

Page 26: ...med for ENHANCE a message with enhance status ENHANCE OFF ENHANCE LOW ENHANCE MEDIUM or ENHANCE HIGH will appear in the message window NOTE When the button is programmed for SHUTTER SPEED UP or DN a m...

Page 27: ...by the key or buttons changes YEAR in order 99 98 00 Month in order 12 11 01 Day in order 31 30 01 Hour in order 23 22 00 Minute in order 59 58 00 and display current setting in the clock display area...

Page 28: ...after word ENHANCE F3 FREEZE Toggles freeze function on and off F4 COPY The copy function is intended to activate image capturing or taping such as video printer VCR etc peripheral equipment please r...

Page 29: ...racter 2 To leave the entering data mode the Esc key shall be pressed after that the cursor shall disappear 3 To reset the patient data the Ctrl and F4 keys shall be pressed simultaneously the data ch...

Page 30: ...window by the key The key shall also move the cursor to the second line the key shall move the cursor up to the first line 2 The protocol of character input shall be similar to described in 3 7 1 NOT...

Page 31: ...A FREEZE KEEP FREEZE REL PICTURE SETTING Copy Mode B COPY SCREEN ADD DATA PIP Location LEFT DN LEFT MID LEFT UP Color Bars ON OFF SCREEN SETTING BUTTON SETTING BUTTON SETTING Button 1 GAIN ENHANCE SH...

Page 32: ...Enzymatic Determent or soap water Use a soft cloth or cotton wool for cleaning Do not use scrubbing material such as a metal brush that might scratch the surface The cleaning procedure should be perfo...

Page 33: ...era head s cable is disconnected from the camera control unit protected by the connector s protecting cap 4 After soaking in disinfecting solution dry with soft cloth in particular any exposed glass w...

Page 34: ...rating life Maintenance services can be obtained from your local representative or from the manufacturer 2 Defective items of equipment are to be serviced and repaired exclusively by persons authorize...

Page 35: ...are correctly set D Dismount the camera from the endoscope and point the camera head towards a light source If the screen becomes brighter check the endoscopic equipment The picture is monochromatic C...