background image

The HART Communicator informs 
the user that the maintenance 
flags have been reset when the 
transmitter has replied to the 
reset command.  The user must 
acknowledge this information by 
pressing OK to continue.

The user is then notified that 
the loop may now be returned 
to automatic control.  After the 
user presses OK the HART 
Communicator returns to the 
Diagnostics screen.

The user must use the left arrow 
to return to the Diag/Service 
screen.

If the user selects Loop test 
from the Diag/Service screen, a 
procedure will begin that will allow 
the user to set the loop current to 
a value desired to perform a loop 
test function.

If the user wishes to specify a loop 
current other than 4mA or 20mA, 
he must select the option Other 
and then press the ENTER key.

The HART Communicator will then 
command the transmitter to set the 
loop current to 4mA and then it will 
dynamically read the value of loop 
current that the transmitter has 
applied.  To return to the previous 
screen, the user must press OK.  
The procedure is exactly the same 
for the 20mA selection option

The user is presented with four 
selections as shown on the 
screen.  The 4mA and 20mA 
selections will set those values 
while selecting Other will require 
the user to input the exact loop 
current that is desired.  The 
procedure will be exited if the user 
selects End.  The example shows 
the user selection 4mA.  The user 
must press ENTER before the 
procedure continues.

When the user selects the 
Loop test option, the HART 
Communicator displays the 
warning message that the loop 
should be removed from automatic 
control during this procedure.

The user must then input the exact 
value of loop current desired.  After 
entering the value, he must press 
ENTER before the procedure 
continues.  For example, the value 
12.00 is input on the screen to the 
left.

The HART Communicator will then 
command the transmitter to set the 
loop current to 12.00mA and then 
it will dynamically read the value 
of loop current that the transmitter 
has applied.  To return to the 
previous screen, the user must 
press OK.

  

To exit the Loop test procedure, 
the user must select the option 
End and then press the ENTER 
key.

  

The HART Communicator 
then issues the command to 
the transmitter to return to the 
measurement mode as it was 
before entering Loop test.

The user is then notified that 
the loop may now be returned 
to automatic control.  After the 
user presses OK the HART 
Communicator returns to the Diag/
Service screen.

If  the  user  selects  Calibration  from 
the  Diag/Service  screen,  a  number 
of  menu  selections  are  available.  
The user must now select the type of 
calibration he desires to perform.  
The Scaled D/A trim will adjust 
the transmitter so that its current 
output at the 4.00mA and 20.00mA 
output values will agree with the 
plant standard current meter.  

10

Summary of Contents for LTM-300 Series

Page 1: ...81 488 0788 FAX 281 488 7080 EMAIL magtech isemagtech com www isemagtech com LTM 300 Series Magnetostrictive Level Transmitter Instruction Operation Manual 3902 Magnolia Road Pearland Texas 77584 Jan...

Page 2: ...r Installation 3 3 1 Standalone Transmitter Installation 3 3 2 Recommended Wiring Single Loop 4 Section 4 Speci cations Page 4 0 Transmitter Electrical Speci cations 4 4 1 Transmitter Sensor Tube 4 Se...

Page 3: ...hange and continuously activates the sensor in the tube The electronics in the housing process the changing signal and update an analog 4 20 mA output This analog output is precisely proportional to t...

Page 4: ...e base to resolve sub microsecond timing increments The binary number equivalent to the microseconds of the echo travel time is written to an output D A Converter and subsequently converted to a 4 20...

Page 5: ...ectronicshousingmountedatthebottomofthegageforeasyaccess THIS MUST BE SPECIFIED AT TIME OF PURCHASE Installation Note European The cable entry devices and blanking elements of unused apertures shall b...

Page 6: ...gurations A single output version with only one variable that will output a 4 20 mA signal for level The second con guration is a transmitter tted with a second oat to provide a digital output for in...

Page 7: ...elect Data not required Span Trim Move the oat to the 20mA mark 100 on the probe Using the arrow buttons input the correct distance from the zero point and then Select Level O set Enter the desired va...

Page 8: ...ion Problems 6 2 Magnetic Interface 6 3Troubleshooting Power Supply Problems Never move a magnet in a perpendicular motion from along the sensor tube This will always leave a residual eld in the waveg...

Page 9: ...as this could leave magnetic indentation in the sensor wire The LTM 300 level transmitter has an inductor located inside the bulkhead of the sensor tube During operation of the transmitter this induc...

Page 10: ...d the Alarm type In addition this screen presents the procedure to select the Primary variable PV if the transmitter is con gured with both Level and Interface It also displays the menu that will allo...

Page 11: ...t are not normal Any status items not reported are within the normal range of operation or calibration The HART Communicator issues a command to receive the results of the transmitter diagnostics Duri...

Page 12: ...xited if the user selects End The example shows the user selection 4mA The user must press ENTER before the procedure continues When the user selects the Loop test option the HART Communicator display...

Page 13: ...output to 4 000mA the user is to enter the reference meter reading For example if the meter actually reads 4 234mA that value is input with the keypad The user must press ENTER when the current value...

Page 14: ...matic control After the user presses OK the HART Communicator returns to the Calibration screen The HART Communicator is now at the starting screen for the Zero Level Trim Full Level Trim Starting at...

Page 15: ...now be returned to automatic control After the user presses OK the HART Communicator returns to the Calibration screen The HART Communicator is now at the starting screen for the Full Level Trim Zero...

Page 16: ...ed Once the measurement is stable the procedure will continue without user intervention The HART communicator will then send the command to the transmitter to calculate the new Temperature zero calibr...

Page 17: ...t Value for 4mA and 20mA by Keypad If you desire to change the Level value at which the Current output is 4mA and 20mA when the Level is the selected Primary Variable or change the Interface value at...

Page 18: ...g the edit box to continue with the reranging of the transmitter On returning to the Keypad rerange screen the values shown for LRV and URV are the changes that the user has made The user should check...

Page 19: ...pper range value getting an update of the digital reading from the transmitter or leaving the 20mA range point as it was If the user accepts the measurement value as the 20mA loop current point the HA...

Page 20: ...h is set in the factory when the unit is prepared for shipment The probe length is used as a reference for a number of diagnostics fuctions as well as setting the URL and LRL shown on the Keypad reran...

Reviews: