CHAPTER 6 REMOTE OPERATION
6-6
MODE DIRECT
Character: “m”
Parameter:
mode number
Returns:
none
Description: Sets the Model 2130 to the mode specified by
mode number. Mode number is
defined in the following table.
Example:
Sending the command “m01” will set the Model 2130 to Dual Limits mode.
Mode
Mode number
Normal
“00”
Dual Limits
“01”
Relative
“02”
Table 6.4.4 Mode Numbers
MODE QUERY
Character: “K”
Parameter: none
Returns:
mode
Description: Returns the current mode setting of the Model 2130.
Example:
Sending the MODE QUERY command while the Model 2130 is set to the Normal
mode will return the following string:
NORM<CR><LF>
RANGE
Character: “N”
Parameter: none
Returns:
none
Description: Simulates the front panel RANGE key.
RANGE DIRECT
Character: “n”
Parameter:
range number
Returns:
none
Description: Sets the Model 2130 to the range specified by
range number. Range number is
defined in the following table.
Example:
Sending the command “n03” will set the Model 2130 range to the lowest range
available. Note: the available ranges depend on the Coarse Range Switch
setting, the Coil Resistance parameter and the Calibration Coefficient.
Range
Number
Range
Gain
“03”
lowest
8
“02”
2
nd
lowest
4
“01”
2
nd
highest
2
“00”
highest
1
Table 6.4.2 Range Numbers
Summary of Contents for 2130