CHAPTER 7 SERVICE
7-2
Message
Indication
“Magnetic Instrumentation”
This message is displayed each time power is
applied to the Model 2130.
“Power Supply Error”
Indicates a failure in the Model 2130 ±5V power
supply. When the Model 2130 is being
powered from an external ±9VDC source, this
error will result if the external source voltage is
too low. If this message is displayed while
powering the unit from an AC source, check to
be certain the correct line voltage is being used.
If this message persists, the unit should be
returned to Magnetic Instrumentation, Inc. for
repair.
“ERROR: Flash Mem”
Indicates a failure in the Model 2130 flash
memory. The Model 2130 will function
normally, however, all configuration settings will
be set to default. The unit should be returned
to Magnetic Instrumentation, Inc. for repair.
“CheckSum Error”
Indicates a check sum error in the Model 2130
program memory. The unit should be returned
to Magnetic Instrumentation, Inc. for repair.
Table 7.2.1 Model 2130 Display Messages