Tube fl ushing
Flush the tube from the air pressure source to the air supply unit connectors and each scale connector.
Flushing cleans the tubes, prevents the embedding of foreign objects in the equipment, and is also useful
to check the tubes.
Air pressure source
If a compressor is selected for use, take into consideration the air consumption amount (approx. 30 N /min
per scale). Th
e graph below shows the discharged air amount and rated output as a rough guideline for
selecting a compressor. Although the specifi c graph may vary depending on the compressor model, please
use it as a reference.
0
0
1
00
200
300
400
500
600
700
800
9
00
1
000
1
2
3
4
5
6
7
8
9
1
0
Dis
c
harged
air
amount
and
rated
output
(N
/min)
(PS)
Dis
c
har
ged
air
amount
Rated
output
Fig. 6-3
Summary of Contents for SR27A-AZ Series
Page 3: ...SR27A AZ Series CH22 Series J 1...
Page 4: ...SR27A AZ Series CH22 Series 2 J...
Page 6: ...ii J SR27A AZ Series CH22 Series...
Page 10: ...4 J SR27A AZ Series CH22 Series 1 6 2006 42 EC...
Page 11: ...J 5 SR27A AZ Series CH22 Series 2 SR27A AZ CH22 S N...
Page 12: ...6 J SR27A AZ Series CH22 Series 3 8 3 1 CH22 SR27A M4 M8 M4 B A M8 3 2 3 2 1 3 1 ML 3 2 3 2...
Page 13: ...J 7 SR27A AZ Series CH22 Series 2 3 M8 16 20 N m M4 10 2 7 N m M4 20 2 7 N m 3 3 3 4...
Page 15: ...J 9 SR27A AZ Series CH22 Series 3 3 7 2 1 0 1 MG 0 05 MG 0 0 2 0 05 3 8 mm...
Page 16: ...10 J SR27A AZ Series CH22 Series 2 2 0 1 MG 3 9 3 t 1 0 mm t 1 0 mm 3 10 t 1 0 mm 1 mm mm...
Page 17: ...J 11 SR27A AZ Series CH22 Series 3 3 5 mm 1 0 N m 242 1 0 N m 1 3 2...
Page 27: ...J 21 SR27A AZ Series CH22 Series 9 3...
Page 28: ...22 J SR27A AZ Series CH22 Series...
Page 32: ...ii E SR27A AZ Series CH22 Series...
Page 54: ...22 E SR27A AZ Series CH22 Series...
Page 58: ...ii G SR27A AZ Series CH22 Series...
Page 80: ...22 G SR27A AZ Series CH22 Series...
Page 82: ......
Page 83: ...SR27A AZ Series CH22 Series 1...