X_Home Sensor Fail
•
No response from X-Home sensor
1 Check if the sliders are blocked Check if X-axis can move smoothly Clean sliders and rails
2 Check if the sensor is functioning properly
3 Check sensor block plates position
4 Check if the motor is functioning properly
5 Check PLC X10 terminal – make sure the cable(CBL-0032) is not loose
6 Check motor connection to power (M001), PCB (S001) and Encoder (M002)
*Error 20 can be triggered in case of misuse and rail/slider rusting To clean the rust follow the
directions below:
- Use large amount of water and grease on the rails and in the sliders (make sure grease goes inside),
clean them If the axis is stuck you can push it to move it, but be careful do not push or shake too hard
- it may mess the calibration up
We suggest to use a soft cloth and a lot of water and grease
To clean the bearings use grease gun to inject the grease inside the slider
Make sure the cables and connectors are not getting wet
If there are rusty stains on the rails that don't go out, they can be cleaned with sanding paper, washed
and greased
- Run X- motor test to lubricate the bearings inside If there is still rust on the rails repeat until they are
clean and lubricate them
After cleaning the rust, keep an eye on the machine It will get stuck again if it's not running for a long
time Run motor test daily at the end of operation to make sure rust is not accumulating It is probable
that X-axis movement won't be as smooth as it was before and rails might make some noise
If rust in the bearings cannot be cleaned the and X-axis is still immobilized, rails and sliders will have to
be replaced This is level 3 maintenance so the machine will have to be sent to RBC Bioscience Corp
Taiwan
Summary of Contents for Plus II
Page 1: ...MagCore PlusII Troubleshootingguide Ver 2020 1230 www rbcbioscience com...
Page 8: ...6 www rbcbioscience com PCBConfiguration...
Page 36: ...34 www rbcbioscience com AxisMisplacement AxisDirections...
Page 61: ...59 MagCore PlusII Measureheatervoltage...
Page 72: ...70 www rbcbioscience com 8 GotopageSystemStatusandchecktheProgramVer...