Maintenance-Teaching Position
Back
7 8 9
0
0
0
4 5 6
1 2 3
07/13/2020 09:25:32
Clear
Enter
0
0
0
0
0
0
TIP4 Position
X Axis Motor
Y Axis Motor
Encode
Input Pulse
Input Pulse
Upload
Encoder Value
Download
Encoder Value
TIP3 Position
Encode
━ X-Axis Calibration
1.
Put the Tips (200 μ l) at the W4 in the positions 1 and 16 on the T-rack.
2. Run W4 piston test and observe the distance between real position and correct
position.
3.
X position:
1 mm = 92 Pulse;
Direction: Inward (X axis moves inward, pistons will be outward) → minus;
Outward → plus.
For example: if you want to make the X axis position (T-rack) 1 mm inward, the original
value is 10000, 10000 – 92 = 9908, enter 9908.
Adjust the Input Pulse → tap "Enter" → tap "Upload Encoder Value" → tap "Execute",
repeat until the position is correct.