Use of the hand control
Quick Start
Set up the patient and proceed as below:
Beginning
Keys to press
Display
Remarks
Switch the unit ON
Check if the hand control is not locked
(See page 5).
To choose the movement
The display shows the new movement selected
and the default settings of the upper and lower
limits of this movement.
Successive presses on the keys
or
allows to select different movements.
Start the session with the original parameters of the
movement (default setting).
The value change at the speed of the
movement.
Possible values for each parameter:
1
st
press
2
nd
press
3
rd
press
1
st
press
2
nd
press
3
rd
press
•
Lower
limit
-90° -50° -50° -30° 0° -30° 0° 0° 0°
• Upper limit
90° 90° 140° 60° 180° 225° 90° 180° 70°
• Speed
1 to 9 (from 150° to 440° per minute)
• Force
1 to 6
• Extension Pause
0 to 900 seconds (15 minutes)
• Flexion Pause
0 to 900 seconds (15 minutes)
• Timer
No timer (00H00) to 24H00
Default setting values for each parameter
1
st
press
2
nd
press
3
rd
press
1
st
press
2
nd
press
3
rd
press
• Lower limit
-10°
-20°
0° 0° 0° 0° 0° 0° 0°
•
Upper
limit 10° 20° 60° 40° 90° 50° 40° 90° 40°
• Speed
2
• Force
6
• Extension Pause
0
• Flexion Pause
0
• Timer
0
Please wait
KINETEC
Maestra V01.0
Prona/supination
-10 STOP 5 10
dip ext/fle
0 run 20 40
Prona/supination
-10 STOP 5 10
WRIST EXT/flex
-20 STOP 94 20
WRIST+MP EXT/fle
0 STOP 94 60
Ulnar dev.Add/ab
0 STOP 94 40
Thumb opposition
0 STOP 94 90
Tpm ext/fle
0 STOP 94 50
ext/fle
0 STOP 94 40
Intrin.- ext/fle
0 STOP 94 90
dip ext/fle
0 STOP 94 40