Maestra
EN
11/21
Use of the hand control
How to define the upper and lower movement limits?
• At the start of a session
The MANUAL MODE is a way to set within the tolerance of a patient at the beginning of a session.
Set up the patient and proceed as below:
Beginning
Keys to press
Display
Remarks
Switch the unit on
Check if the hand control is not locked
(See page 5).
To select the MANUAL MODE for upper limits by
continually holding pressure on the key
continuous press
The unit is moving to the upper limit of the
movement.
To set the pain level when reached, immediately press
The new upper value limit of the movement is
recorded.
To select the manual mode for lower limits
continuous press
The unit is moving to the lower limit of the
movement
To set the pain level when reached, immediately press
The new lower value limit of the movement is
recorded.
To start the session with the new movement limits
The angle display changes with current
movement.
• During the session
This function, The BY-PASS MODE is a way to address the pain threshold of a patient during a session.
IMPORTANT: Can be used only when the machine RUN.
Beginning
Keys to press
Display
Remarks
The unit is running
Check if the hand control is not locked
(See page 5).
To select the BY-PASS MODE
continuous press
The unit exceeds the recorded upper limit.
To set the new pain level when reached, immediately
press
The new upper value limit of the movement is
recorded.
To select the BY-PASS mode for lower limits
continuous press
The unit is moving to the lower limit of the
movement.
To set the new pain level when reached, immediately
press
The new lower value limit of the movement is
recorded.
Continue the session with the new movement limits.
The angle display changes with current
movement.
Comments:
• The values shown in the ‘Display’ column are examples.
Prona/supination
-5 Manual 5 15
Prona/supination
-5 Manual 5 15
KINETEC
maestra V01.0
Please wait
Prona/supination
-10 STOP 5 10
Prona/supination
-10 Manual 15 5
Prona/supination
-10 Manual 15 15
Prona/supination
-10 Manual 5 15
Prona/supination
-10 STOP 5 10
Prona/supination
-30 Bypas 30 40
Prona/supination
-30 RUN 10 40
Prona/supination
-10 BYPAS 40 15
Prona/supination
-10 Bypas 40 40
Prona/supination
-30 Bypas 30 40