Maestra 8080 User Manual Download Page 11

Maestra

 

 

   

EN   

 

10/21 

 

Use of the hand control 

 

Using Programs 

 

The KINETEC Maestra allows you to store up to 16 programs, including the type of movement, ROM, speed, load, pauses and timer. 

 

The original parameter values of the program are empty. 

These values can be modified and recorded at any time (see ‘How to enter a program’ page 10). 

 

 

Beginning 

Keys to press 

Display 

Remarks 

To Stop the Unit 

 

 

Check if the hand control is not locked  
(See page 5). 

To access the program mode 

 

 

The program number blinks. 

To change the program if necessary 

or 

 

The new program number blinks. 

To exit and validate the selected program 

 

 

The current parameters have been recorded in 
program 3. 

To exit without validation of selected program 

 

 

Back to the starting parameters. 

Start the unit 

 

 

The value change at the speed of the 
movement. 

 

 

Reading the values of a program: example SPEED 

 

Beginning 

Keys to press 

Display 

Remarks 

To Stop the Unit 

 

 

Check if the hand control is not locked  
(See page 5). 

To access the program mode 

 

 

The program number blinks. 

To change the program if necessary 

or 

 

To change the program if necessary. 

To read the speed value 

 

 

Displays the speed value. 

After 5 seconds or after pressing on another key 

 

 

 

To exit and validate the selected program 

 

 

The current parameters have been recorded in 
program 3. 

Start the unit 

 

 

The value change at the speed of the 
movement. 

 

 

 

Comments: 

• The values shown in the ‘Display’ column are examples. They actually depend on the stored programs. 

• The current movement parameters can be changed while using that program but no data will be stored in the original program.  

    See the programming mode (p 8) to modify programs. 

Prona/supination

-10     STOP    5   10

WRIST       EXT/flex

-30  prg nr2  30 

Thumb opposition

0 prg nr 5    30 

Thumb opposition

0  stop      10    30

Thumb opposition

0  stop       20   30

Prona/supination

-10     STOP    5   10

WRIST       EXT/flex

25    STOP     20    90

Prona/supination

-10 prg nr 4  50 

THUMB OPPOSITION

0  prg nr 3      50

 

THUMB OPPOSITION

   SPEED :      2

 

WRIST       EXT/flex

25    STOP     20    90

THUMB OPPOSITION

 0  STOP    20     50

 

THUMB OPPOSITION

 0   RUN    20     50

 

Summary of Contents for 8080

Page 1: ...User manual Before use please read this document Kinetec reserves the right to effect technical modifications EN 467896276 08 2011 MAESTRA...

Page 2: ...____________________________________________ 15 Thumb opposition __________________________________________________ 16 Complete fist _____________________________________________________ 17 Intrinsic...

Page 3: ...ulotomies arthrolysis and tenolysis for post traumatic stiffness of MCP PIP and SIP joints Synovectomies Clinical Benefits Breaks the cycle of trauma inflammation and the loss of range of motion Preve...

Page 4: ...Spare parts 10 Miscellaneous textiles 11 Ulnar deviation mobilization accessories 12 Power supply cable transport position 13 Hand control 14 Liquid crystal display 2 lines of 16 characters 15 Select...

Page 5: ...not touch the moving parts while the unit is running pinching risk Electrical Connection KINETEC Maestra is a type B class I device Before connecting the device to the power supply check that the main...

Page 6: ...hine as appropriate Simultaneously press the and keys to lock the hand control The display reads LOCK you can not change the parameters if you try the display reads LOCK To unlock the hand control sim...

Page 7: ...press Lower limit 90 50 50 30 0 30 0 0 0 Upper limit 90 90 140 60 180 225 90 180 70 Speed 1 to 9 from 150 to 440 per minute Force 1 to 6 Extension Pause 0 to 900 seconds 15 minutes Flexion Pause 0 to...

Page 8: ...pause in extension or flexion limit Successive presses on this key selects the pause at the extension or flexion limit To change the pause if necessary or The new pause value blinks To validate the ne...

Page 9: ...record the program 10 Then The program 10 has been recorded and the display indicates the next program so you can change another program OR To cancel the program The program 10 has been cancelled and...

Page 10: ...omments When a program has been deleted the display shows Program table Timer Pause on lower limit Pause on upper limit Force Speed Upper Limit Lower limit Movement type Program number 1 2 3 4 5 6 7 8...

Page 11: ...eck if the hand control is not locked See page 5 To access the program mode The program number blinks To change the program if necessary or To change the program if necessary To read the speed value D...

Page 12: ...address the pain threshold of a patient during a session IMPORTANT Can be used only when the machine RUN Beginning Keys to press Display Remarks The unit is running Check if the hand control is not lo...

Page 13: ...lexion The joints also permit the forearm support to be tilted up or down However if this is done the angle of tilt should be taken into account when setting the range of motion on the display Setting...

Page 14: ...er fit individual hand lengths and polar radii depending on the fingers or thumb to be mobilized the guide will be oriented as follows by unscrewing the nut 8 and lifting the guide 7 in position A or...

Page 15: ...pination Choice of the Pronation Supination movement Press and adjust your parameters see page 7 Or select a program see page 10 Extension Flexion of the wrist The KINETEC Maestra provides a motion fr...

Page 16: ...ion from 50 to 140 Choice of the 2nd movement of the wrist Press 2 times and adjust your parameters see page 7 Or select a program see page 10 Ulnar deviation The KINETEC Maestra provides a motion fro...

Page 17: ...de with the screw 1 Remove the guide 2 from Set up the extension 3 of guide Set up the driving mechanism the driving mechanism Set up the guide 2 on the motor hub see p 12 for more details Set up the...

Page 18: ...o 225 Choice of the complete fist movement Press and adjust your parameters see page 7 Or select a program see page 10 Intrinsic The KINETEC Maestra provides a motion from 0 from 90 Choice of the intr...

Page 19: ...ment Press 2 times and adjust your parameters see page 7 Or select a program see page 10 Selective mobilization of the DIP joint The KINETEC Maestra provides a motion from 0 to 70 Choice of the Select...

Page 20: ...100 240 V Frequency 50 60 Hz Power consumption 50 VA Device of type B class I IP 20 Fuse T 750mA 250V 6 3x32mm KINETEC order 4610007434 Environment Storage transport conditions Temperature 40 to 160...

Page 21: ...t of the component or components found to be defective KINETEC guarantees its continuous passive motion systems for 2 years against all defects of manufacture from the date of purchase by the consumer...

Page 22: ...______________ Technician ____________________________________________ _________________________________________ Running time see page 18 _________________________________ ____________________________...

Reviews: