Manual
GA-60
To delete a report highlight it and press
Delete
function key, you will then see the
confirmation screen like the one below:
Drawing 33.
Internal memory erasing process confirmation screen.
Pressing the
All
function key will erase all the reports stored in the internal
memory of the analyser.
5.3.6.4. Gas conditioner
Drawing 34.
Gas conditioner screen.
This screen shows settings for the external gas conditioning unit
PGD-100
that
can work in cooperation with the
GA-60
.
(+)
and
(-)
function keys allow changing settings for the conditioning unit (gas
dryer). Setting available include: pump efficiency, heated hose temperature and the
temperature of cooling unit.
The column Is displays the set values read from the conditioner while Set shows
setting entered by the user manually through the analyser OR the conditioner.
34