Manual
GA-60
Attention!
Fresh air must be provided to the analyser through the analyser's
probe throughout the whole process of zeroing.
If an
auto-start
option is chosen in the saving options, and an SD/MMC card is in
the card port, the third screen visible will include request for clients number for which all
following measurements will be carried (you can find more information about auto-start in
chapter 5.3.3.3)
5.2. Preparation for measurements
Before you start the analyser you should set up a complete measuring module:
Connect a probe pipe with a probe holder
Connect a gas hose to the analyser (both connectors, gas and electric)
Connect the analyser to the mains power supply (if analyser is to work from
the battery this step can be skipped)
Start the analyser with the switch on the front panel (right side of the panel,
close to the handle)
Then a process of zeroing the sensors will start and after about 2 minutes the
analyser is ready to work.
Attention!
Fresh air must be provided to the analyser through the analyser's
probe throughout the whole process of zeroing.
5.3. Measurements screens
5.3.1. Measurements parameters screens
Measurements parameters screen is the first that you can see when the analyser
is ready to work. You can also get to this view from Results screens
or
Menu screens.
11