Manual
GA-60
Drawing 6.
Measurements parameters screen.
From this place you can change analysers measurements parameters. Use arrow
keys up/down to move between different settings. Keys left/right will allow changing
chosen parameters value (that is highlighted with a small arrow on the left side of the
screen). You can also use function key
change
to increase selected value.
You can change below mentioned parameters:
Averaging time
– time interval for which all measurement results will be
recalculated averaged and displayed.
GA-60
reads measurements results
every 2 seconds and stores the last 90 measurements in its inner memory.
Analyser can display value averaged for chosen time from these stored
results. You can set the averaging time to one of these values: 2s, 6s, 10s,
20s, 30s, 60s, 120s, 180s.
Attention!
This operation will change the averaging time globally. If the
averaged results are being saved, all of them will be stored with the
new time set.
Measurement time XL
– it is a time of single XL measurement.
GA-60
measures for the set time, averages the result and stores it as a report in its
inner memory or on a SD/MMC card. XL time can be set to: 10s, 20s, 30s,
1min, 2min, 5min, 10min, 15min, 20min, 30min.
NO in NOx
– allows setting value of percentage content of NO in NOx (nitric
. If the analyser is equipped with NO
2
sensor,
NO in NOx setting will not be visible.
12