U5098 Series Ultrasonic Sensors Manual
099-MF179 - Electrical Connections/Configuration Modes
Page. 12 of 40
www.madisonco.com
U5098 -
electrical connections /
conÞ
guration modes
1
2
3
USB/RS485
POWER
SUPPLY
HUMIDITY INFILTRATIONS
To avoid the
humidity inÞltration inside
the
housing is
recommended:
- to
use a cable
with a 0.2”-0.4” (
5-10mm) outer diameter and fully tighten the M16 cable gland for electrical
connections
-
to fully tighten the
cap
- to
position the cable so that
it
forms a downward curve
at
the M16 output; in this
way
the condensation and/or
the
rain water will
tend to
drip from the curve
bottom
7-CONFIGURATION
MODES
The U5098 have 2
conÞguration/calibration
modes:
- via digital communication: via MODBUS RTU
by
PC
- via
2 on board
buttons
DIGITAL COMMUNICATIONS CONNECTION
U5098 MODBUS RTU PC connection
1) U5098 with
MODBUS
RTU
communication protocol
2)
USB/RS485 interface
module
3)
MODBUS
RTU
communication S/W, for
U5098
transmitter
With this software is
possible:
-
to connect the
U5098
transmitters
in
MODBUS RTU network by selecting the UID
address
-
to read
on
your PC monitor all measures
in
reading and U5098 operation
data
- to
program all U5098 conÞguration parameters
-
to store the measurement readings
and the
operating status
of U5098