made for movement Innowalk Troubleshooting Manual Download Page 8

 

Troubleshooting Guide 

Innowalk 2.gen 

 

Document code: 

T36 

Made by: 

FGB 

Date: 

25.03.21 

Revision:

 

Approved by: 

AL/KU 

Date

25.03.21 

Page: 

8 of 12 pages 

 

 

Ski loose from drive unit arms 

15 

The ski should be tight relative 
to the rear module.  

Open the round side cover by unscrewing the screws. Tighten the nut 
holding the skis with a 22mm wrench:  
On the left side: It is a left-handed thread on the left side, turn counter-
clockwise to tighten. 
On the right side: Normal threads on this side. Turn clockwise to tighten. 

Contact 
service 
personnel 

Walking function stops when 

applying load 

16 

Spasm control  

NEVER adjust the spasm control without the proper technical and 
therapeutistic insight and NEVER without training. Only representatives 
approved by Made for Movement are allowed to adjust the spasm control, 
as errorous use/adjustment here can injure the user. 

Contact 
service 
personnel. 
The service 
personnel will 
assess if the 
error is 
caused by 
hardware 
error or 
errorous 
adjustment, 
and initiate 
proper 
actions to 
address the 
issue. Do NOT 
use the 
Innowalk 
again until it 
has been 
examined by 
service 
personnel. 

 

 

Summary of Contents for Innowalk

Page 1: ...umns below Illustrations of the topics and affected parts are found on the following page after the troubleshooting tables in the pages below where numbers on the illustrations are reflected onto the...

Page 2: ...heck power cable Visually inspect the power cable for damage Unplug and insert the power cable firmly into the machine and into the AC power socket Continue troubleshooting steps 4 5 If broken or dama...

Page 3: ...Troubleshooting Guide Innowalk 2 gen Document code T36 Made by FGB Date 25 03 21 Revision A Approved by AL KU Date 25 03 21 Page 3 of 12 pages 1 2 3 AC Power outlet socket 4 5...

Page 4: ...le on the remote control change remote control If broken socket on the machine contact service personnel 7 Check remote control Try all functions on the machine by pushing all buttons on the remote co...

Page 5: ...ecute the homing sequence Important to hold 5 seconds after the actuator stops moving to finalize the sequence Contact service personnel to replace actuator or electronic box Seat height actuator does...

Page 6: ...Troubleshooting Guide Innowalk 2 gen Document code T36 Made by FGB Date 25 03 21 Revision A Approved by AL KU Date 25 03 21 Page 6 of 12 pages 6 7 8 9 and 10...

Page 7: ...ontact service personnel Knee unfolding mechanism 13 Knee joint mechanism If a user is applying extensive force into the calf bows the knee opening mechanism can be hard impossible to open then pull t...

Page 8: ...ormal threads on this side Turn clockwise to tighten Contact service personnel Walking function stops when applying load 16 Spasm control NEVER adjust the spasm control without the proper technical an...

Page 9: ...Troubleshooting Guide Innowalk 2 gen Document code T36 Made by FGB Date 25 03 21 Revision A Approved by AL KU Date 25 03 21 Page 9 of 12 pages 11 12 12 Swinger on top of motor cover seen from above 12...

Page 10: ...Troubleshooting Guide Innowalk 2 gen Document code T36 Made by FGB Date 25 03 21 Revision A Approved by AL KU Date 25 03 21 Page 10 of 12 pages 13 14...

Page 11: ...Troubleshooting Guide Innowalk 2 gen Document code T36 Made by FGB Date 25 03 21 Revision A Approved by AL KU Date 25 03 21 Page 11 of 12 pages 15 16...

Page 12: ...by AL KU Date 25 03 21 Page 12 of 12 pages CONTACT INFORMATION Document history Revision Description of change Date Initials A First edition 25 03 2021 FGB B Main office email info madeformovement co...

Reviews: