made for movement Innowalk Troubleshooting Manual Download Page 2

 

Troubleshooting Guide 

Innowalk 2.gen 

 

Document code: 

T36 

Made by: 

FGB 

Date: 

25.03.21 

Revision:

 

Approved by: 

AL/KU 

Date

25.03.21 

Page: 

2 of 12 pages 

 

 

ELECTRONICS SIDE PANEL 

Topic 

No. 

Parts to check 

Actions 

Still not fixed? 

Product will not 

power up 

Check main power ON/OFF 
switch 

Turn the switch ON 

Continue 
troubleshooting 
steps 2-5. 

Check emergency stop button 

Rotate the red mushroom button in the direction of the arrows to release 
the emergency stop button. 

Continue 
troubleshooting 
steps 3-5. 
If faulty emergency 
stop button. 
Contact service 
personnel to 
replace button. 

Check power cable 

Visually inspect the power cable for damage. Unplug and insert the power 
cable firmly into the machine and into the AC power socket. 

Continue 
troubleshooting 
steps 4-5. 
If broken or 
damaged power 
cable and/or 
socket on machine. 
Replace cable 
and/or power 
socket. 

Check fuse 

Unscrew the fuse holder and visually inspect the fuse. If broken, insert new 
fuse according to specification printed on sticker on the switching panel.  
(If the Innowalk works after changing the fuse, and there has been no 
external power influences towards the Innowalk (eg.thunderstorms) please 
contact your Made for Movement representative for a control of the 
product) 

Continue 
troubleshooting 
step 5. 
 

Check power connection 

Try different AC power outlet socket 

Contact service 
personnel 

 

Summary of Contents for Innowalk

Page 1: ...umns below Illustrations of the topics and affected parts are found on the following page after the troubleshooting tables in the pages below where numbers on the illustrations are reflected onto the...

Page 2: ...heck power cable Visually inspect the power cable for damage Unplug and insert the power cable firmly into the machine and into the AC power socket Continue troubleshooting steps 4 5 If broken or dama...

Page 3: ...Troubleshooting Guide Innowalk 2 gen Document code T36 Made by FGB Date 25 03 21 Revision A Approved by AL KU Date 25 03 21 Page 3 of 12 pages 1 2 3 AC Power outlet socket 4 5...

Page 4: ...le on the remote control change remote control If broken socket on the machine contact service personnel 7 Check remote control Try all functions on the machine by pushing all buttons on the remote co...

Page 5: ...ecute the homing sequence Important to hold 5 seconds after the actuator stops moving to finalize the sequence Contact service personnel to replace actuator or electronic box Seat height actuator does...

Page 6: ...Troubleshooting Guide Innowalk 2 gen Document code T36 Made by FGB Date 25 03 21 Revision A Approved by AL KU Date 25 03 21 Page 6 of 12 pages 6 7 8 9 and 10...

Page 7: ...ontact service personnel Knee unfolding mechanism 13 Knee joint mechanism If a user is applying extensive force into the calf bows the knee opening mechanism can be hard impossible to open then pull t...

Page 8: ...ormal threads on this side Turn clockwise to tighten Contact service personnel Walking function stops when applying load 16 Spasm control NEVER adjust the spasm control without the proper technical an...

Page 9: ...Troubleshooting Guide Innowalk 2 gen Document code T36 Made by FGB Date 25 03 21 Revision A Approved by AL KU Date 25 03 21 Page 9 of 12 pages 11 12 12 Swinger on top of motor cover seen from above 12...

Page 10: ...Troubleshooting Guide Innowalk 2 gen Document code T36 Made by FGB Date 25 03 21 Revision A Approved by AL KU Date 25 03 21 Page 10 of 12 pages 13 14...

Page 11: ...Troubleshooting Guide Innowalk 2 gen Document code T36 Made by FGB Date 25 03 21 Revision A Approved by AL KU Date 25 03 21 Page 11 of 12 pages 15 16...

Page 12: ...by AL KU Date 25 03 21 Page 12 of 12 pages CONTACT INFORMATION Document history Revision Description of change Date Initials A First edition 25 03 2021 FGB B Main office email info madeformovement co...

Reviews: