PM100 Bump Test and Calibration Station
REV – 1.0
[34-2900-0513-2 ]
16
|
P a g e
software shows the available configuration options for PM
R
100
R
devices. All configuration settings can be set to be
simultaneously updated during each bump test or calibration (see Section 5.3.1.2 Enable/Disable Function).
5.3.2.1
PM100 Configuration Options
Each configuration option can be enabled/disabled using the checkbox on the left side. Disabling the option will not
disable the feature in the PM
R
100
R
monitor. It will only prevent the Bump-Calibration Station from changing the
existing setting in each PM
R
100
R
monitor.
Note
: Although the software will perform a “Read” operation when opening, If the device is not in place, the “Read”
operation may need to be performed once the software is opened to display the configuration of the device.
User ID
– The User ID field can be used to add a custom message.
Calibration Interval (Days)
– The Calibration reminder can be configured from n/a (Off) to 365 days.
Bump Interval (Days)
– The Bump Test reminder can be configured from n/a (Off) to 365 days.
Self-Test Interval (hours)
– The Self-Test reminder can be configured to n/a (Off) or between 8 hours to 20
hours.
Low Alarm
– This is the first (low for O
R
2
R
) set point that activates the monitor’s alarms.
High Alarm
– This is the second (high for O
R
2
R
) set point that activates the monitor’s alarms.
Display
– Choose whether the monitor display shows the current gas reading or the remaining sensor life.
5.3.2.2
Gas Concentration