PM100 Bump Test and Calibration Station
REV – 1.0
[34-2900-0513-2 ]
14
|
P a g e
Docking Time Auto Sync
– If checked, the time on the Bump-Calibration Station will be synced to the computer
time. If unchecked, the time can be manually adjusted within the Time tab.
Location
A custom location can be added up to 12 alphanumeric characters. This can be useful to keep data from
multiple Bump-Calibration Stations organized.
Gas Expiration
– Displays the expiration date of the installed gas cylinder. Update this field when a new gas cylinder
is installed. Gas expiration date can be found on the gas cylinder.
Gas Bottle Lot #
- Displays the lot number of the installed gas cylinder. The gas cylinder lot number can be found on
the gas cylinder. Update this field when a new gas cylinder is installed.
Info Read/Write
– Press
Read
to download and display the current settings of the Bump-Calibration Station.
Pressing Read will discard any unsaved changes in the Station tab. Press
Write
to upload the settings currently
displayed on the software to the Bump-Calibration Station. No changes will be saved to the Bump-Calibration
Station until the
Write
button is pressed.
5.3.1.2
Enable/Disable Function
When the Bump-Calibration Station performs a bump test or calibration on a PM
R
100
R
device, it can optionally perform
other maintenance tasks simultaneously.
Calibrate
– If the Calibrate box is unchecked, the calibrate function of the Bump-Calibration Station will be disabled.
With the calibrate function disabled, pressing the Calibrate button will not calibrate the monitor, but it will update
configuration or firmware if the option is enabled. The power LED will turn orange for 5 seconds, then return to
green.
Bump Tests
– If the Bump Tests box is unchecked, the Bump Test function of the Bump-Calibration Station will be
disabled. With the Bump Test function disabled, pressing the Bump Test button will not test the monitor, but it will
update configuration or firmware if the option is enabled. The power LED will turn orange for 5 seconds, then return
to green.
Update PM100 Config
– If this box is checked, the Bump-Calibration Station will update the PM
R
100
R
configuration
settings (See Section 5.3.2.1 PM100 Configuration Options) when the Bump Test or Calibration button is pressed.
Update PM100 Firmware
– If this box is checked, the Bump-Calibration Station will upgrade the PM
R
100
R
firmware
when the Bump Test or Calibration button is pressed. It is not necessary to upgrade the PM
R
100
R
firmware every time
the monitor is tested. It is only necessary when a newer version of the firmware is available. To update the firmware
file saved on the Bump-Calibration Station, see Section 5.3.2.4 PM100 Firmware Upgrade.
Bump Result Decision
– The dropdown list is selectable for “Beep”, “Concentration”, or “Beep + Concentration”.
The “Beep” option will run the bump test until the alarm condition is met and the audible alarm is detected.