PM100 Bump Test and Calibration Station
REV – 1.0
[34-2900-0513-2 ]
17
|
P a g e
This field allows you to enter the gas concentration(s) of the test gas cylinder installed in the Bump-Calibration
Station. This is necessary for proper testing and calibration of PM
R
100
R
monitors. There are fields available for H2S, CO,
and O2.
5.3.2.3
PM100 Option Info Read/Write
IMPORTANT: Any changes made to the PM100 tab will not be saved to the Bump-Calibration Station until the
“Write” button is clicked.
To write the data from the PM100 tab to the Bump-Calibration Station, press
Write
.
To read the current PM
R
100
R
configuration options saved in the Bump-Calibration Station, press
Read
.
NOTE
: Pressing
Read
will cause any unsaved changes on the PM100 tab to be lost.
5.3.2.4
PM100 Firmware Upgrade
The Bump-Calibration Station can be used to upgrade the PM
R
100
R
firmware. The Bump-Calibration Station can keep a
copy of the PM
R
100
R
firmware saved on its USB Disk. PM
R
100
R
monitors require a firmware file that is specific to the
monitor’s gas type. If there is a new firmware version, the file can be found at https://macurco.com/downloads/.
To update the PM
R
100
R
firmware file,
1.
Press Browse to open the file explorer.
2.
Navigate to the PM
R
100
R
firmware file.
3.
Press Open.
4.
Press
Write
. The PM
R
100
R
firmware file will be saved to the USB Disk of the Bump-Calibration Station.