Manual
Instructions for use
DSPS
-
Prominent
SOLUTIONS IN RADIOTHERAPY
5
19092022V2IFU113000
MacroMedics ®
™
4. Baseplate set-up
SECURING THE DSPS-
PROMINENT
BASEPLATE TO THE COUCHTOP
I.
Position the two dedicated DSPS-
Prominent
couchstrips on the couchtop.
For more information, refer to IFU147100 - CouchStrips and Sidepieces.
For use in MR, use the glass fiber, MR Safe 3-Pin couchstrip.
II.
Position the DSPS-
Prominent
above the couchtop. Then, align the slots at the bottom of the
baseplate with the two couchstrips and carefully place the baseplate on the couchstrips.
5. Mask set-up
• Follow your department’s patient positioning protocol.
•
For DSPS treatments, there should be approximately one finger width in
between the cradle rim and the top on the patient’s head.
• When using a mask on a patient with long hair, place the hair in
a ponytail on top of the patient’s head.
• It is strongly recommended to use the MacroMedics mouldroom
tool to ensure the smoothest possible mask set-up process.
• Screw the mouldroom tool to the DSPS-
Prominent
frame.
This tool will help to determine the correct head depth.
Mouldroom Tool
Point of
attachment