Manual
Instructions for use
DSPS-Prominent
SOLUTIONS IN RADIOTHERAPY
8
06042023V4IFU113000
MacroMedics ®
®
9.
Push the white frame slowly towards the black frame. Once both frames touch, secure the mask
in place by pushing the sliding lock pawls into locked position. Be careful not to pinch the
patient’s ears or skin around the neck between the frames.
10. Now secure the ends of the upper frame to the cradle as well, just as described in step 6 of
chapter 5.1, by locking the second rotatable lock pawl -
Only applicable for the head, neck
and shoulder masks.
11.
Swing the top part of the Mouldroom Tool above the patient’s head again to check if the chin
still fi ts underneath it.
12.
Use both hands to mould the structures of the face like the nose bone, orbitae, chin, and
forehead.
13.
After moulding the facial region of the mask, proceed with moulding the mask around the neck
and shoulders -
Only applicable for the head, neck and shoulder masks.
14.
To speed up the setting process, you can use cold packs or cold, wet towels to cool the mask.
Cure and cool the mask for at least 5 minutes.
15.
Remove the Mouldroom Tool and label the mask(s) with the patient’s name/ identifi er.
Upper frame
now locked
Upper frame
unlocked
GF DSPS-Prominent Baseplate with moulded
Head, neck & shoulder Facial and Occipital Masks
CF DSPS-Prominent Cradle with moulded
Head Only Facial and Occipital Masks