MS574 0210 0007
Page 19
4.4.
INSTALLING THE HARDWARE
4.4.1. INTRODUCTION
Electro static discharge does inflict irreparable damage to electronics! Before installing
or opening the unit, the installer has to discharge himself by touching a well-grounded
object.
This unit must be installed in accordance with the EMC guidelines (Electro Magnetic
Compatibility).
Fig. 7: Grounding aluminum enclosure
FOR INSTALLATION, PAY EMPHATIC ATTENTION TO:
Separate cable glands with effective IP67 (NEMA4X) seals for all wires.
Unused cable entries: ensure that you fit IP67 (NEMA4X) plugs to maintain rating.
A reliable ground connection for both the sensor, and if applicable, for the metal casing. (above)
An effective screened cable for the input signal, and grounding of it’s screen to the “
┴
“ terminal
or at the sensor itself, whichever is appropriate to the application.
4.4.2.
TERMINAL CONNECTORS WITH POWER SUPPLY
The following terminal connectors are available:
5
POWER SUPPLY
UNIT
6
1
2
SENSOR SIGNAL
PULSE INPUT
SIGNAL
3
+
+
+
10
POWER SUPPLY
BACKLIGHT
+
SENSOR
SUPPLY
4
9
Fig. 8: Overview of terminal connectors ER