6
Spectrophotometer
NANOCOLOR
®
Advance 1.1, 10.2021
Specification
Description
Wavelength
reproducability:
± 0.1 nm
Scan speed:
1 complete scan < 3 min
Spectral bandwidth:
< 4 nm
Photometric range:
± 3.0 Abs
Photometric accuracy:
0.003 Abs at 0.0–0.5 Abs; 1 % at 0,5–2,0 Abs
Photometric linearity:
< 0.5 % at 0.5–2.0 Abs;
≤ 1 % at > 2 Abs with neutral glass filters at 546 nm
Stray light:
< 0.5 %
Measurement modes:
More than 250 preprogrammed tests and special
methods; 100 free programmable methods; Absorbance;
Transmission; Factor; Kinetics, scan; nephelometric turbidity
measurement
Turbidity measurement:
Nephelometric turbidity measurement (16 mm and 24 mm)
1–1000 NTU
Cuvette slot:
Test tubes 16 mm and 24 mm OD;
Rectangular cuvettes 10 mm, 20 mm, 40 mm and 50 mm
Battery:
Lithium iron phosphate battery; approx. 1000 measurements
or 8 h runtime (depending on
energy settings)
Data memory:
1000 measurement results, 1000 IQC results, 100 scans;
GLP‑conform
Display:
Backlit display
Operation:
Barcode technology; icon‑based menu guidance; capacitive
touch screen
Languages:
DE / EN / FR / ES / NL / IT / HU / PL / CZ / PT‑Br / TK / BG / RO /
DK
External light:
Insensitive; open cuvette slot
Interfaces:
LAN (CAT 6; only use shielded cable of max. 20 m length)
2 × USB (Host), 1 × USB (Function)
Weight:
3.8 kg
Operating conditions:
10–40 °C, up to 80 % rel. humidity (not condensing),
max. 3000 m
Protection class according
to IEC 60529:
IP 68 (Device is dust and waterproof)
IP6X
≜
Dust proof
IPX8
≜
Protection against temporary submersion (30
minutes at 1 m water depth)