12
Spectrophotometer
NANOCOLOR
®
Advance 1.1, 10.2021
Step 6: Result screen
Fig. 8: Display of result
The end of the measurement is announced by
an acoustic signal. The measurement result is
shown on the display (Fig. 8). In addition to
the measurement result, information on wave‑
length, absorbance and interfering turbidity is
also given, depending on the settings.
If several results are available for one method,
they are displayed next to each other in the
form of tabs.
NOTICE
If the result of the measurement is outside the measuring range of the used test
kit, the result is displayed with "< measuring range" or "> measuring range".
Below the result an estimated value is displayed in square brackets.
Step 7: Enter sample information
Fig. 9: Sample information
Below the measurement result are the icons
for entering the sample information. The
information comment, sample location and
dilution are set here ex works.
You can change this selection and enter
the sample information by clicking on
the
‑icon. Use the sliders to set whether
the corresponding sample information is
displayed (Fig. 9).
The entered information is saved together
with the measurement result.