After a PCB Enclosure Assembly has been installed or reset, it must be programmed.
7. Programming a replacement/reset control panel
The display is flashing «068» (US – 110 V version «027»)
1.
Press the button
to drive to the desired lower end position (or to the under block position).
The system moves downwards at half speed. Upward movement is disabled.
2.
Press the buttons
1
(plus) and
2
(minus) to set the current height of the work surface on the
display. (in cm, US - 110 V version in inch)
3.
To confirm, press
S
(Save).
After confirmation the display changes to «088» (US - 110 V version «035») (still flashing).
4.
Press the button
to drive to the desired upper end position (or to the upper block position).
5.
Press the buttons
1
(plus) and
2
(minus) to set the current height of the work surface on the
display. (in cm, US - 110 V version in inch)
6.
To confirm, press
S
(Save).
After confirmation the height is displayed (no more flashing) and the initial operation is completed.
NOTE
The control box automatically offsets the end positions by one motor turn.
Depending on the combination of systems (hydraulic transmission ratio), the system stops
its movement 3 mm, 5 mm or 10.5 mm before the defined end position.