background image

47LR

P. 5 / 4

EM-9503-B Rev.

SETTING PROCEDURE

POWER ON

Measuring Mode

Hold down Alarm/    or Init/    button 

for 

1 second

Hold down Init/    button

for 

3 seconds

Measuring

■ 

GENERAL SETTING FLOWCHART

Measuring stopped

*

3

Mode transition

*

1. Preset time can be specified with “Transition Time to Lockout Setting Mode” in Advanced Setting Mode.

*

2. Alarm Setting Mode is locked with no-alarm-output type. 

*

3. The last measured values or status are held for DC and alarm outputs while the measuring is stopped.

  (Except the analog output adjustments during the loop test and the initial setting)

*

1. Optional settings for parameters in a broken-line section.

  Alarm Setting Mode is locked with no-alarm-output type. 

Initial Setting Mode

Alarm Setting Mode

Advanced Setting Mode

Hold down Alarm/    or Init/   

button for 

1 second

Hold down Alarm/    or Init/    button 

for 

1 second

Hold down Max/Min + Alarm/    buttons at once

for a preset time duration

*

1

 

Hold down Alarm/   + Init/    buttons at once

for 

3 seconds

POWER ON

*

1

 

INITIAL SETTING FLOWCHART

Initial Setting Complete

Measuring Started

Hold down Init/    button for 3 seconds or more to move on to Initial Setting Mode.

Choose an input type and display range values.

Hold down Alarm/    button for 3 seconds or more to move on to Alarm Setting Mode.

Go through the alarm settings.

Hold down Alarm/   + Init/    buttons at once for 3 seconds or more 

to move on to Advanced Setting Mode.  Go through the advanced settings.

Hold down Alarm/    button or Init/    button for 1 second or more

to go back to Measuring Mode.

If necessary, hold down Max/Min Alarm/    buttons at once

for a preset time duration to enter Lockout Setting Mode.

Go through the lockout settings.

Lockout Setting Mode

Hold down Alarm/    or Init/    button 

for 

1 second

Hold down Alarm/    button

for 

3 seconds

Loop Test Output Mode

Hold down Alarm/    or Init/    button 

for 

1 second

Hold down Alarm/   + Init/    + Shift buttons at once

for 

5 seconds

*

2

Summary of Contents for 47LR

Page 1: ...icate it POWER INPUT RATING OPERATIONAL RANGE Locate the power input rating marked on the product and confirm its operational range as indicated below 100 240V AC rating 85 264V 50 60 Hz approx 6 5VA...

Page 2: ...alarm setting mode or to shift through setting items in each setting mode 6 Init button Used to move on to the initial setting mode or to shift through setting items in each setting mode 7 Shift butt...

Page 3: ...must be in place to hold the meter Do not remove it 1 Insert the unit into the panel cutout 2 Push the mounting brackets into the grooves on both sides of the rear module until they hit the panel s r...

Page 4: ...b NC Ha NO 17 15 16 20 18 19 Lc LLc LLa La Hc HHc Ha HHa LL OUTPUT L OUTPUT H OUTPUT L OUTPUT H OUTPUT HH OUTPUT 9 10 POWER U V ALARM SUFFIX CODE 2 SPDT contact 2 points 9 10 POWER U V ALARM SUFFIX CO...

Page 5: ...Hold down Max Min Alarm buttons at once for a preset time duration 1 Hold down Alarm Init buttons at once for 3 seconds POWER ON 1 INITIAL SETTING FLOWCHART Initial Setting Complete Measuring Started...

Page 6: ...t button shifts one to the previous Changing parameters Pressing Shift button while one of the parameter settings is indicated on the display shifts the panel meter into the setting standby mode The d...

Page 7: ...eters 1999 9999 UP DOWN 1 Hold down Init button for 3 seconds Hold down Alarm or Init button for 1 second Shift Shift Alarm Alarm Alarm Alarm Alarm Alarm Alarm Shift Shift Alarm Alarm Initial Setting...

Page 8: ...decimal fraction is specified Temperature values must be re set if the decimal point position is shifted Temperature input offset setting 4 Pt 100 JIS 97 IEC 5 Pt 50 JIS 81 1 decimal place No decimal...

Page 9: ...Init Alarm Alarm Init Alarm Alarm Setting Mode Alarm Shift L Setpoint L Deadband L ON Delay Time 1999 9999 0000 9999 00 99 Shift Shift Alarm Init Alarm Alarm Init Alarm Alarm Init Alarm Init L Coil a...

Page 10: ...ix code 2 is specified A B B B B B B Init HH ON Delay Time 00 99 Shift Alarm Init Alarm Alarm Init Alarm Init HH Coil at Alarm RYEN RYDN Shift Init Init Init Alarm Init Alarm Shift HH Setpoint 1999 99...

Page 11: ...m H Coil energized at alarm H Fnc H Coil de energized at alarm D P H Fnc H P Fnc HH setpoint HH deadband hysteresis HH ON delay time HH ON delay time seconds HH Deadband hysteresis value HH Setpoint v...

Page 12: ...ghtness level 4 Brightness level 5 bright Brightness level 3 Brightness level 2 No moving averaging INDICATORS FUNCTION DEFAULT VALUE DISPLAY Fnc D P Fnc D P Fnc D P Fnc Transition time to Lockout Set...

Page 13: ...for a preset time 1 1 Preset time can be specified with Transition Time to Lockout Setting Mode in Advanced Setting Mode A Init Init Init Init Set value Setting Parameters A Alarm setting lockout Lock...

Page 14: ...If the unit does not recover its function after the initialization repairing in the factory may be required Increase decrease the input signal until it is back within the measuring range Check the inp...

Reviews: