Safety Precautions
1.
Heed all warning labels. No warning label should ever be removed from the unit.
2.
Connect the chiller to a properly fused disconnect box in compliance with the National
Electric Code (NFPA-70) as well as state and local codes for American usage, or national
and local codes for European usage. Maximum fuses must not exceed the maximum rating
found on the serial tag on the electrical box. To reduce the risk of electric shock:
!
Disconnect electric power before opening the electrical box of the unit.
!
Do not operate with electrical box door open.
!
Refer servicing of electrical box components to qualified/certified personnel.
!
Do not operate equipment with damaged electrical power cords.
!
Turn off the unit and disconnect electrical power before servicing or moving.
!
Properly ground the unit.
3.
The servicing of the refrigeration system or components should be performed only by
properly certified refrigeration technicians.
4.
Coolant lines, filters, and other components which connect to the unit must be capable of
withstanding the maximum pressure that the pump in the chiller can deliver at the maximum
expected temperature.
5.
Do not remove any warning labels.
6.
When operating a pump with a magnetic coupling, pay special attention to the handling of
permanent magnetic parts. Due to potential interference from magnets, keep a safe
distance from pace makers and other devices subject to magnetic interference.
7.
Fluids can be extremely hot or cold. Use appropriate precautions when contact with fluids
is possible.