7. COUNTING SCALE OPERATION
PROCEDURE
1) Prepare a lot of counting sample units. The no. of
counting sample will be 10 PCs, 20 PCs, 50 PCs or 100
PCs.
The more PCs of counting sample will be better.
* No load to put on the platform.
2) Use the finger to hold the " CAL. Button ( function A
Button ) " ( 3-3, Fig. 1 ), at the same time set the
" Power Switch " ( 3-6, Fig. 1 ) to the " On " position.
Until one beeper sound is present, then release the
finger away from " function A Button ".
3) a. The bottom left corner of LCD will show the
" COUNT " marker, at the same time the LCD will
show no. of sample units from 10, 20, 50, 100 in
sequence & changed per second.
b. Put the counting sample units ( For example 10 PCs,
20 PCs, 50 PCs or 100 PCs ) on the platform.
c. Until the display reach the no. of sample units
exactly, for example " 50 ", push the " function A
Button " ( 3-3, Fig. 1 ) once a while. The display
will flash twice and freeze the sample counting no.,
then take away the sample units. Now the
balance is ready for counting.