Page
38
of
60
Chapter 5: Camera Operation
NOTE: Unless specifically noted for one or the other, the following information is true for both the LT-300 SD
and the LT-300 HD digital video colposcopes.
WARNING:
•
At any point during the use of the colposcope, if any liquid gets splashed onto the outer casing or the lens of
the colposcope, immediately stop using the device and wipe the liquid off. Please follow the instructions set
forth in chapter 6 of the manual under Cleaning and Maintenance.
•
At any point during the use of the colposcope, if any liquid gets splashed onto the colposcope and gets
inside the casing, please turn off the colposcope right away and call your technician or the manufacturer.
The device should be checked by an authorized technician to make certain it is working accurately before
being put back to use.
5.1 Preliminary Check List
Before using the colposcope, check the following to ensure safe operation:
•
Make sure there is no mechanical damage.
•
Inspect all the cables, parts and accessories and make certain they have all been installed accurately and
securely.
•
Inspect all parts and make sure they are in good condition.
•
Inspect the external condition of the colposcope to make sure it is clean and that there is no damage to the
enclosure.
•
Make sure there are no liquids on the outer surface, lens, or inside the colposcope.
•
Check to make sure the connection between the camera, camera mount, and vertical stand is tight, secure
and stable.
If any damage is found or any necessary parts or accessories are missing, do not use the colposcope.
Please contact the hospital engineer or manufacturer-authorized service center immediately.
5.2 Function Buttons
Function buttons on the camera:
Diagram 5.2-1
1
Power: Turns colposcope ON/OFF
2
Green/White Filter: Toggles between the preset green filter and the white filter.
3
Freeze: Freezes the live feed image on the screen.