background image

Guidance and manufacturer’s declaration – electromagnetic immunity

The LuminaNRG Evolution (Model: LuminaNRG Evolution) is intended for use in the 

electromagnetic environment specified below. The customer or the user of the 

LuminaNRG Evolution (Model: LuminaNRG Evolution) should assure that it is used in 

such an environment.

Immunity 

test

IEC 60601

test level

Compliance

level

Electromagnetic environment – guidance

Conducted

RF

IEC

61000-4-6

Radiated 

RF

IEC

61000-4-3

3 Vrms

150 kHz to 

80MHz

3 V/m

80 MHz to 

2.5GHz

Not 

Applicable

3 V/m

Portable and mobile RF communications 

equipment should be used no closer to any 

part of the LuminaNRG Evolution  (Model: 

LuminaNRG Evolution), including cables, 

than the recommended separation distance 

calculated from the equation applicable to 

the frequency of the transmitter. Recom-

mended separation distance

Where P is the maximum output power 

rating of the transmitter in watts (W) 

according to the transmitter manufacturer 

and d is the recommended separation 

Distance in meters (m). 

Field strengths from fixed RF transmitters, 

as determined by an electromagnetic site 

survey, 

a. should be less than the compliance level 

in each frequency range. 

b. Interference may occur in the vicinity of 

Equipment marked with the following 

symbol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propaga-

tion is affected by absorption and reflection from structures, objects and people.

a. Field strengths from fixed transmitters, such as base stations for radio 

(cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM 

radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. 

To assess the electromagnetic environment due to fixed RF transmitters, an 

electromagnetic site survey should be considered. If the measured field strength in 

the location in which the LuminaNRG Evolution (Model: LuminaNRG Evolution) is 

used exceeds the applicable RF compliance level above, the LuminaNRG Evolution 

(Model: LuminaNRG Evolution) should be observed to verify normal operation. If 

abnormal performance is observed, additional measures may be necessary, such 

as re-orienting or relocating the LuminaNRG Evolution (Model: LuminaNRG 

Evolution).

b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 

3 V/m.

Recommended separation distances between portable and mobile RF 

communications equipment and the Model LuminaNRG

The LuminaNRG Evolution (Model: LuminaNRG Evolution) is intended for use in an 

electromagnetic environment in which radiated RF disturbances are control RGB. The 

customer or the user of the LuminaNRG Evolution (Model: LuminaNRG Evolution) can 

help prevent electromagnetic interference by maintaining a minimum distance between 

portable and mobile RF communications equipment (transmitters) and the LuminaNRG 

Evolution (Model: LuminaNRG Evolution) as recommended below, according to the 

maximum output power of the communications equipment.

Rated maximum 

output power of 

transmitter (W)

Separation distance according to frequency of transmitter (m)

150 kHz to 

80 MHz

80 MHz to 

800 MHz

800 MHz to 

2.5 GHz

0.01 

         0.12                         0.12 

                           0.23

0.1 

         0.38                         0.38 

                           0.73

         1.2                         1.2 

                            2.3

10 

          3.8                         3.8 

                            7.3

100 

          12                          12 

                            23

For transmitters rated at a maximum output power not listed above, the 
recommended separation distance d in meters (m) can be estimated using the 
equation applicable to the frequency of the transmitter, where P is the maximum 
output power rating of the transmitter in watts (W) according to the transmitter 
manufacturer.
NOTE1 At 80 MHz and 800 MHz, the separation distance for the higher frequency 
range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic 
propagation is affected by absorption and reflection from structures, objects and 
people.

17

18

Summary of Contents for Evolution

Page 1: ...EMS LED Facial Toning Therapy ...

Page 2: ...non invasive combination of light and micro current technology LuminaNRG targets fine lines deep wrinkles enlarges pores and your blemishes will shrink and fade Painless and easy to use LuminaNRG will quickly help you achieve clear beautiful healthy looking skin all from the comfort of your home Clear your vanity of clutter since LuminaNRG can be used on all skin types to diminish the look of line...

Page 3: ...ED Treatment LuminaNRG has 2 lights red blue allows you to harness the power of all 2 colors simultaneously This is perfect for those with more complicated complexions and multiple skin concerns NOTE LED lights ilLuminaNRGte when contact with the skin is established EMS Treatment The EMS Electronic Muscle Stimulation function can help rejuvenate the look and feel of wrinkles aging or sagging skin ...

Page 4: ... activator to prime skin for the EMS treatment and allows your device to glide easily and comfortably overskin This gentle formula contains Organic Electronic Muscle Stimulation EMS EMS can help rejuvenate the look and feel of wrinkled aging or sagging skin While using EMS an electric current stimulates specific muscles under the surface of the skin Your face will immediately feel more toned while...

Page 5: ...color Treatment twice a week for maintenance Place into base for charging until the next treatment To best protect the delicate light technology always store in its holding base EMS Treatment Instructions Generously apply LuminaNRG Micro Current Activator Gel to all treatment areas of your choice including cheeks chin T zone and forehead This formula enhances the effect of the EMS function Applyin...

Page 6: ...t deep mario nette lines running from below your nostrils to your chin area as well as tiny wrinkles and frown lines near your lips Never use LuminaNRG on the lips or eyes LuminaNRG Micro Current Activator Gel LuminaNRG comes packaged with a gel primer which acts as a micro current activator to prime skin for the EMS treatment and allows your device to glide easily and comfortable over skin This g...

Page 7: ...ealth If this is not the case for you please consult your doctor before use If using the device caused any discomfort discontinue use immediate ly If you suspect that you had a negative reaction from using the device consult your doctor immediately If you have suspected or diagnosed heart disease you should follow precautions recommended by your doctor Caution should be used in the following situa...

Page 8: ...Package the item securely and return it prepaid insured along with proof of purchase to Heat In A Click LLC Warranty Repair Department 1975 tigertail blvd Dania Florida 33034 USA To insure prompt repair provide complete legible name address and phone number information RMA number and a note indicating the nature of the product defect and a copy of the original invoice issued for purchase of the un...

Page 9: ...urposes Guidance and manufacturer s declaration electromagnetic immunity The LuminaNRG Evolution Model LuminaNRG Evolution is intended for use in the electromagnetic environment specified below The customer or the user of the LuminaNRG Evolution Model LuminaNRG Evolution should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environ...

Page 10: ...netic site survey should be considered If the measured field strength in the location in which the LuminaNRG Evolution Model LuminaNRG Evolution is used exceeds the applicable RF compliance level above the LuminaNRG Evolution Model LuminaNRG Evolution should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as re orienting or reloc...

Page 11: ...requency 8 33Hz Maximum Output Voltage 10 0 254V 500Ω 1 04V 2kΩ Maximum Output Current 10 0 508mA 500Ω 0 52mA 2kΩ Net Charge per pulse 60 98uC 500Ω Maximum Current Density 0 627mA cm2 500Ω The Minimum Electrode Size 0 81cm Time Range 5 minutes Photon Mode Specification Wavelengths Red RGB Light 620 730 nm Wavelength Blue RGB Light 430 450 nm Wavelength Heat Temperature Range 41 2 Energy Red RGB Li...

Page 12: ...the manufacturer s serial number Batch code 9 Manufacturer Information Company Name Heat In A Click LLC Address 1975 tigertail blvd Dania Florida 33004 USA Phone 1 954 518 9777 Fax No 1 954 320 7984 Designed in USA Copyright 2016 2 Face Face Evolution Utility and Design patents pending LuminaNRG Evolution This product is brought to you by Heat In A Click LLC Address 1975 tigertail blvd Dania Flori...

Reviews: