background image

Safety, EMC & Biocompatibility

1) This device is Class II equipment with type BF applied part. The light treatment 

attachment and electrical muscle stimulation treatment attachment were 

considered as one type BF applied part. It complies with Medical Electrical Safety 

Standards (IEC 60601-1, IEC 60601-2-10).

2) This device is also complied to Medical EMC Standard (IEC 60601-1-2).

3) All the user directly contracting materials for main device housing and output 

contacts in this device are biocompatible for its intended use. They are complied 

with biocompatibility standards ISO 10993-5 (Cytotoxicity) and ISO 10993-10 

(Irritation and Sensitization).

Guidance and manufacturer’s declaration – electromagnetic emissions

The LuminaNRG Evolution (Model: LuminaNRG Evolution) is intended for use in 

the electromagnetic environment specified below. The customer or the user of the 

LuminaNRG Evolution (Model: LuminaNRG Evolution) should assure that it is used 

in such an environment.

Emissions    Compliance    Electromagnetic environment-- guidance

RF 

emissions

CISPR 11

Group 1

The LuminaNRG Evolution (Model: LuminaNRG 

Evolution) uses RF energy only for its internal function. 

Therefore, its RF emissions are very low and are not 

likely to cause any interference in nearby electronic 

equipment.

RF 

emissions

CISPR 11

Class B

Class A

Harmonic

emissions

IEC 

61000-3-2

Voltage

fluctuations/

flicker 

emissions

IEC 

61000-3-3

Not 

Applicable

The LuminaNRG Evolution(Model: LuminaNRG 

Evolution) is suitable for use in all establishments, 

including domestic establishments and those directly 

connected to the public low-voltage power supply 

network that supplies buildings used for domestic 

purposes.

Guidance and manufacturer’s declaration – electromagnetic immunity

The LuminaNRG Evolution (Model: LuminaNRG Evolution) is intended for use in the 

electromagnetic environment specified below. The customer or the user of the 

LuminaNRG Evolution (Model: LuminaNRG Evolution) should assure that it is used in 

such an environment.

Immunity 

test

IEC 60601

test level

Compliance 

level

Electromagnetic 

environment

--guidance

Electrostatic

discharge 

(ESD) IEC 

61000-4-2

±6 kV 

contact

±8 kV air

±6 kV 

contact

±8 kV air

Floors should be wood, concrete or 

ceramic tile. If floors are covered with 

synthetic material, the relative humidity 

should be at least 30 %.

Electrical 

fast transient

/burst IEC 

61000-4-4

±2 kV for 

powersupply 

lines

±1 kV for

input/output 

lines

Not 

Applicable

Not 

Applicable

Not 

Applicable

Mains power quality should be that of 

a typical commercial or hospital 

environment

Surge

IEC 

61000-4-5

± 1 kV 

line(s) and

neutral

Mains power quality should be that of 

a typical commercial or hospital 

environment.

Voltage dips,

short

interruptions 

andvoltage

variations

on power 

supply input 

lines IEC 

61000-4-11

<5 % UT

(>95 % dip 

in UT)for 0,5 

cycle

40 % UT(60 

% dip in UT)

for 5 cycles

70 % UT(30 

% dip in UT)

for 25 cycles

<5 % UT

(>95 % dip 

in UT)for 5s

Mains power quality should be that of 

a typical commercial or hospital 

environment. If the user of the 

LuminaNRG Evolution (Model: 

LuminaNRG Evolution) requires 

continued operation during power 

mains interruptions, it is recommend-

ed that the LuminaNRG Evolution 

(Model: LuminaNRG Evolution) be 

powered from an uninterruptible 

power supply or a battery.

Power 

frequency

(50/60 Hz)

magnetic 

fieldIEC 

61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields 

should be at levels characteristic of a 

typical location in a typical commercial 

or hospital environment.

NOTE U

T

 is the a.c. mains voltage prior to application of the test level

15

16

Summary of Contents for Evolution

Page 1: ...EMS LED Facial Toning Therapy ...

Page 2: ...non invasive combination of light and micro current technology LuminaNRG targets fine lines deep wrinkles enlarges pores and your blemishes will shrink and fade Painless and easy to use LuminaNRG will quickly help you achieve clear beautiful healthy looking skin all from the comfort of your home Clear your vanity of clutter since LuminaNRG can be used on all skin types to diminish the look of line...

Page 3: ...ED Treatment LuminaNRG has 2 lights red blue allows you to harness the power of all 2 colors simultaneously This is perfect for those with more complicated complexions and multiple skin concerns NOTE LED lights ilLuminaNRGte when contact with the skin is established EMS Treatment The EMS Electronic Muscle Stimulation function can help rejuvenate the look and feel of wrinkles aging or sagging skin ...

Page 4: ... activator to prime skin for the EMS treatment and allows your device to glide easily and comfortably overskin This gentle formula contains Organic Electronic Muscle Stimulation EMS EMS can help rejuvenate the look and feel of wrinkled aging or sagging skin While using EMS an electric current stimulates specific muscles under the surface of the skin Your face will immediately feel more toned while...

Page 5: ...color Treatment twice a week for maintenance Place into base for charging until the next treatment To best protect the delicate light technology always store in its holding base EMS Treatment Instructions Generously apply LuminaNRG Micro Current Activator Gel to all treatment areas of your choice including cheeks chin T zone and forehead This formula enhances the effect of the EMS function Applyin...

Page 6: ...t deep mario nette lines running from below your nostrils to your chin area as well as tiny wrinkles and frown lines near your lips Never use LuminaNRG on the lips or eyes LuminaNRG Micro Current Activator Gel LuminaNRG comes packaged with a gel primer which acts as a micro current activator to prime skin for the EMS treatment and allows your device to glide easily and comfortable over skin This g...

Page 7: ...ealth If this is not the case for you please consult your doctor before use If using the device caused any discomfort discontinue use immediate ly If you suspect that you had a negative reaction from using the device consult your doctor immediately If you have suspected or diagnosed heart disease you should follow precautions recommended by your doctor Caution should be used in the following situa...

Page 8: ...Package the item securely and return it prepaid insured along with proof of purchase to Heat In A Click LLC Warranty Repair Department 1975 tigertail blvd Dania Florida 33034 USA To insure prompt repair provide complete legible name address and phone number information RMA number and a note indicating the nature of the product defect and a copy of the original invoice issued for purchase of the un...

Page 9: ...urposes Guidance and manufacturer s declaration electromagnetic immunity The LuminaNRG Evolution Model LuminaNRG Evolution is intended for use in the electromagnetic environment specified below The customer or the user of the LuminaNRG Evolution Model LuminaNRG Evolution should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environ...

Page 10: ...netic site survey should be considered If the measured field strength in the location in which the LuminaNRG Evolution Model LuminaNRG Evolution is used exceeds the applicable RF compliance level above the LuminaNRG Evolution Model LuminaNRG Evolution should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as re orienting or reloc...

Page 11: ...requency 8 33Hz Maximum Output Voltage 10 0 254V 500Ω 1 04V 2kΩ Maximum Output Current 10 0 508mA 500Ω 0 52mA 2kΩ Net Charge per pulse 60 98uC 500Ω Maximum Current Density 0 627mA cm2 500Ω The Minimum Electrode Size 0 81cm Time Range 5 minutes Photon Mode Specification Wavelengths Red RGB Light 620 730 nm Wavelength Blue RGB Light 430 450 nm Wavelength Heat Temperature Range 41 2 Energy Red RGB Li...

Page 12: ...the manufacturer s serial number Batch code 9 Manufacturer Information Company Name Heat In A Click LLC Address 1975 tigertail blvd Dania Florida 33004 USA Phone 1 954 518 9777 Fax No 1 954 320 7984 Designed in USA Copyright 2016 2 Face Face Evolution Utility and Design patents pending LuminaNRG Evolution This product is brought to you by Heat In A Click LLC Address 1975 tigertail blvd Dania Flori...

Reviews: